![Critikon DINAMAP 8100 Operation Manual Download Page 55](http://html1.mh-extra.com/html/critikon/dinamap-8100/dinamap-8100_operation-manual_2686191055.webp)
6-6
MAINTENANCE
Continued
Figure 6-4. Remove Fuse From Fuse Holder
Replace the fuse and fuse holder as shown in Figure 6-5.
Make sure that the arrows on the outward facing ends of the
fuse holders are pointing in the same direction as the
arrows on the inside of the cover plate. Snap down the
cover plate.
Figure 6-5. Replace AC Line Fuse and Fuse Holder
Replacement of Battery Fuse
Refer to qualified service personnel since placement of the