![Critikon DINAMAP 8100 Operation Manual Download Page 24](http://html1.mh-extra.com/html/critikon/dinamap-8100/dinamap-8100_operation-manual_2686191024.webp)
3-8
TABLE 3-1 MODEL 8100 FRONT PANEL CONTROLS AND
INDICATORS — Continued
I T E M N A M E
F U N C T I O N
15
LOW
LIMIT
This grey momentary pushbutton switch
displays the current low alarm limit that
you select (SELECT switch). If you push
and hold this switch, you will cause the
display to step through the entire range of
low limit settings (settings overlapping the
high limit will not be displayed.) The new
low limit alarm setting will be the setting
that is displayed at the time you release
the LOW LIMIT switch.
16
SILENCE This grey momentary pushbutton switch
with LED indicator alternately mutes and
enables the audio alarm. When pressed
(SILENCE ON) the LED indicator lights to
indicate that the audio portion of the
alarms have been silenced. When an
alarm condition is detected the LED will
flash but no audio alarm will sound.
Press again to enable audio alarm.
17
HIGH
LIMIT
This grey momentary pushbutton switch
displays the current high alarm limit that
you select (SELECT switch). If you push
and hold this switch, you will cause the
display to step through the entire range of
high limit settings (settings overlapping
the low limit will not be displayed). The
new high limit alarm setting will be the
setting that is displayed at the time you
release the HIGH LIMIT switch.