![Critikon DINAMAP 8100 Operation Manual Download Page 22](http://html1.mh-extra.com/html/critikon/dinamap-8100/dinamap-8100_operation-manual_2686191022.webp)
3-6
TABLE 3-1 MODEL 8100 FRONT PANEL CONTROLS AND
INDICATORS — Continued
I T EM N A M E
F U N C T I O N
7
CANCEL
Cont.
Visual and audible alarms may be
canceled during wait time and the
displays will maintain the last determined
values.
8
SET
This grey momentary pushbutton switch
increments the MINUTES display when in
auto mode. The MINUTES increments
are 1, 2, 3, 4, 5,10, 15, 20, 30, 45, 60, and
90 minutes.
Holding the SET switch in for 4 seconds
while momentarily pressing the ON
switch places the monitor in calibrate
mode, where it will remain for 3 minutes.
Holding the SET switch for 20 seconds
while momentarily pressing the ON
switch will cause the monitor to remain in
calibrate mode indefinitely.
9
CYCLE
MINUTES
This 2-digit red LED display shows the
cycle time (in minutes) when the monitor
is placed into automatic mode. The
display is normally blanked when in
manual mode. In either mode, the
elapsed time since the last data update
(up to 90 minutes) is flashed in the
CYCLE MINUTES display.
When auto mode is first entered, the
display indicates a default cycle time of
three minutes. This cycle time can be
changed by using the SET switch.
Periodically, the elapsed time of the
current cycle is displayed briefly to
indicate the amount of time elapsed since
the last data update.