3-10
TABLE 3-1 MODEL 8100 FRONT PANEL CONTROLS AND
INDICATORS — Continued
I T EM N A M E
F U N C T I O N
19
AUTO/
MANUAL
Cont.
This white momentary pushbutton switch
with adjacent LED indicators controls and
indicates the mode of operation for the
monitor. Pressing this switch changes
the operating mode as indicated by the
associated green LED.
Manual mode is entered when power is
first applied to the monitor. In manual
mode, a single determination is made
each time and only when the START
switch is pressed.
In the auto mode, one determination is
initiated immediately. Subsequent
determinations occur at the end of the
cycle time shown in the CYCLE
MINUTES display (Item 9).
Determinations may be initiated at any
time by pressing the START switch
(during wait time). The cycle timer starts
when auto mode is first entered and
restarts at the beginning of each
automatic or manually started
determination.
20
PULSE
bpm
This 3-digit yellow LED display shows
pulse rate. In addition, this LED display:
■
shows the alarm codes during certain
alarm conditions. Refer to Section 5,
Troubleshooting.
■
shows pulse rate alarm limits (see
Item 18, SELECT).