LPT-M - Operation Manual
8.2 Calibration Adapter Clip “Cal Clip”
(p/n: CET-SGC)
To calibrate an LPT-M with a factory installed splash guard (Option -S), attach the Cal Clip around
the splash guar
d. The small barb hose fi tting accommodates standard or Tefl on tubing of 1/8”
(3.175 mm) ID and 3/16” (4.762 mm) ID. The Cal Clip
must be removed during normal
operation
or else the gas readings will not be accurate.
8.3 Magnetic Wand
(p/n: CET-MW)
The magnetic wand is used for accessing menu
options and program functions without opening
the enclosure. Use to engage the magnetic
sensors on the side of the LPT-M.
Lifts
226.7 g (½ lb) solid steel
Size
66.6 mm x 6.35 mm (2 5/8” X 1/4”) Hexagon
8.4 Metal Protective Guard
(p/n: SCS-8000-RSG)
The metal protective guard (p/n:
SCS-8000-RSG
) is made of heavy gauge, galvanized metal and
helps protect against abrasive damage, theft and vandalism to the transmitters. The 16-gauge
galvanized steel has 13 mm (½ in) square openings in the front to allow gas and air to fl ow through