LPT-M - Operation Manual
scroll through the menu and select the channel for which you want to test the
gas reading response.
Press ENTER and use the ARROW keys to scroll through the menu and choose
TEST. Press ENTER and use the ARROW key to increase or decrease the value
and the ENTER key to move to the next digit.
NOTE:
You will only be able to enter a value that falls within the factory
configured range of the sensor. If you try to enter a number outside that limit,
the value will return to a valid value.
When finished, press ENTER to accept and Exit the screen. The gas level
reading for that channel will be immediately sent to the controller. This test
reading will override any actual gas reading for that channel for the duration of
the test.
To stop the test, press ENTER to Exit. If you do not exit the test, the test will