PREVENTIVE MAINTENANCE

 

Some  preventive  procedures  are  necessary  for  the  best  functioning  and  durability  of  your  autoclave,  the 

preventive maintenance corresponds to the 

fulfilment of all procedures listed

 below.

 

Use only distilled water

;

 

Keep the autoclave clean. For the 

anodized trays/aluminum chambers or stainless steel chambers

, use only a 

soft, 

non-abrasive

 sponge with biodegradable neutral detergent and distilled water. To remove the foam use a cloth 

that does not shed. Finish the cleaning with another cloth and alcohol 70% or peracetic acid at 1%. 

ATTENTION!

 The 

use of other materials and/or products may scratch or damage these parts;

 

The outside cleaning of the autoclave must be done 

daily

 using a soft cloth and biodegradable neutral detergent. 

Next, clean it thoroughly with a cloth and alcohol 70% or peracetic acid at 1%. The handle must be also cleaned the 
same way before removing the material from the autoclave, after the sterilization;

 

Clean the door gasket 

weekly 

with a clean cloth that does not shed, dampened in warm water. Replace the door 

gasket of your autoclave

 annually;

 

Perform the preventive maintenance on the VSPF valve 

annually

 (it consists of replacing the internal safety valve that 

comprehends the seal and pin), replace the air filter of the VSPF valve 

monthly

.

 

Replace the thermal paste of the heating element 

annually

;

 

The component “lid” (Item 2, page 9) must be replaced each 10 years;

 

Replace the internal hoses 

annually

;

 

Biological tests, air filter replacement and preventive maintenance:

 

The autoclave Vitale Plus 12/21 was programmed to alert the user/operator about some very important sterilization 

procedures:

 

 

- A biological test will be requested 

every 7 days

, the display will show the message: 

PERFORM BIOLOGICAL TEST

For  further  information  about  how  to  perform  the  biological  test  in  your  autoclave,  visit  our  site, 
www.cristofoli.com and go to your product support tab;

 

 

- The autoclave will request the replacement of the air filter 

every 30 dias

, the display will show the message: 

change air filter

. See “

How to Proceed When Service is Needed 

” (Page 24);

 

 

- The cleaning cycle will be requested 

every 30 days

 with the message: 

PERFORM cleaning. see manual

;

 

 

-  The  equipment  will  request  the  preventive  maintenance 

once  a  year

  by  showing  the  message: 

PERFORM 

Prevent

.

 maintenance

. See “

How to Proceed When Service is Needed 

” (Page 24);

 

Attention!

 In order to avoid incrustation, the aluminum chambers of the autoclaves have been anodized. Hence, it 

is forbidden to use any anticrust product to perform the internal cleaning. This type of product will damage the 
chamber of the autoclave.

2

3

1

5

6

7

4

8

9

10

19

CRISTÓFOLI

A U T O C L

A V E S

I N S T R U C T I O N    M A N U A L

Summary of Contents for Vitale Plus 12

Page 1: ...uce its environmental impacts and continuous training of its employees achieving this way a sustainable profitability and the maximization of the company s value Rev 2 Develop innovative solutions to...

Page 2: ......

Page 3: ...rist foli s Biosafety Consultant who provides a valuable assistance in the coordination of Crist foli s Biosafety Project complementary products research training of our employees representatives and...

Page 4: ...ymbols Introduction Important Safety Information Installation Instructions Autoclave Components Identification Safety Devices Cycles and Messages Safety Notes How to Use the Autoclave Vitale Plus Alti...

Page 5: ...nding Protection Terminal Fusible Autoclavable Date of manufacture Boas Pr ticas de Fabrica o 4 P B F LOT SN Fragile Handle with care Dangerous Electrical Tension Serial Number Batch code Keep away fr...

Page 6: ...ation we recommend extra attention to this manual Autoclave sterilization although quicker and more efficient requires totally different procedures and handling while still keeping a simple operation...

Page 7: ...y kind of use of this equipment other than the ones described in this manual SPECIAL MEASURES AND OBSERVATIONS DURING THE USE OF THE AUTOCLAVE Always make sure the door handle is properly locked befor...

Page 8: ...he use of an exclusive circuit breaker for the outlet where you intend to install your autoclave If after following all the specifications the electrical current is still oscillating contact your loca...

Page 9: ...of the external steam exit 8 Fig 1A Fig 3 WATER V CUT DO NOT USE THE LID KEEP THE HOSE ABOVE THE WATER LEVEL 5 liters 1 3 gallons Table 2 X NEVER CUT THE CENTRAL PIN OFF Fig 1 EXTERNAL STEAM EXIT VSPF...

Page 10: ...use interference in the cycle EXTERNAL STEAM EXIT Located in the back part of the autoclave Fig 1 page 8 it has a diameter of 5 16 for the connection of the discharge hose which is then connected to t...

Page 11: ...7 Door Gasket Fig 5 Display Start Key Mode Key Cancel Key Lid Panel Keyboard Handle Fig 4 CRIST FOLI A U T O C L A V E S I N S T R U C T I O N M A N U A L Fig 8 Fig 10 Fig 9 250 ml 50 ml 200 ml 150 m...

Page 12: ...onents Identification Figs 6 and 7 page 10 VSPF VALVE Valve that works as an anti vacuum system patent pending FUSE Safety device which purpose is to protect the building electrical wiring against pea...

Page 13: ...an no perform cleaning see manual perform biological test cooling 00 00 T 000 C p 0 00 cooling 00 00 cooling T 000 C power outage subpressure detected cancelled by user overpressure detected temp x pr...

Page 14: ...ymbol 14 appears in some places on the autoclave This means those items require special attention and that the user must observe their references in the Instruction Manual provided with the equipment...

Page 15: ...re for the new cycle is reached the autoclave will resume operation showing heating up When the solenoid valve shuts down the display will show the temperature and pressure rising gradually When the a...

Page 16: ...the CANCEL key to return to the program selection mode ATTENTION Even after the beeps that indicate the conclusion of the cycle the contents of the chamber will still be hot Never touch the internal...

Page 17: ...me ALTITUDE CALIBRATION In order for the autoclave to work properly it is necessary to perform the altitude calibration procedure Calibrate the altitude of the autoclave as shown below for this proced...

Page 18: ...valve closed this will prevent the liquid from boiling over inside the containers when the liquids sterilization cycle was selected If the operator needs to interrupt cancel the heating sterilization...

Page 19: ...ature of drained water Total heat in joules transmitted in one hour Proper working temperature range Proper working altitude 12 liters Aluminum 19 4 kg including components Stainless steel 21 2 kg inc...

Page 20: ...eplace the air filter of the VSPF valve monthly Replace the thermal paste of the heating element annually The component lid Item 2 page 9 must be replaced each 10 years Replace the internal hoses annu...

Page 21: ...hamber Preventive cleaning stainless steel chamber Door gasket cleaning Door gasket replacement Preventive Maintenance of the VSPF valve and replacement of the internal safety valve Air filter replace...

Page 22: ...pose of the label located in the back of the equipment is to identify the autoclave technical data ATTENTION The removal of the identification label and or any other labels stickers affixed to the pro...

Page 23: ...on See How to Proceed when Service is Needed Page 24 THE AUTOCLAVE DOES NOT SWITCH ON POSSIBLE CAUSES SOLUTION The operator connected the autoclave to the power source but didn t press the Start key N...

Page 24: ...g the sterilization of 26 envelopes at a time 10 x 23 cm containing 6 instruments each Remember to leave some space between the envelopes to allow good steam circulation and optimize drying THE AUTOCL...

Page 25: ...ive Maintenance Item 7 Page 19 Crist foli Equipamentos de Biosseguran a Ltda makes clear its concern already demonstrated by the implementation of the Environmental Management System according to stan...

Page 26: ...rcuits fire etc Damage caused by humidity excessive exposure to sun light and salinity Sinister theft or robbery Improper use installation of the equipment and or plugging it into an incorrect voltage...

Reviews: