Fig.19

ATTENTION!

When opening the autoclave for cooling, observe

the correct position of the door in the picture. The autoclave

Vitale Plus 

dries with the door closed

, its opening is only

necessary to speed up the cooling process of the 

instruments/articles sterilized).

Upper view

CRISTÓFOLI

A U T O C L

A V E S

I N S T R U C T I O N    M A N U A L

When the depressurization is over, the drying stage will begin automatically, the display will show the message 

drying

 

and the temperature of the chamber. 

The autoclave Vitale Plus dries with the door closed

. At the end of the drying 

process, the autoclave will beep 3 times and the display will show the message 

cooling

. At this moment, the operator 

must  open  autoclave  door  and  leave  it  ajar  to  accelerate  the  cooling  process  of  the  sterilized  articles.  Within  5 
minutes  the  autoclave  will  beep  continuously  and  the  display  will  show  the  message 

cycle  concluded

,  only  then, 

remove the sterilized articles. Press the 

CANCEL

 key to return to the program selection mode.

ATTENTION!

 Even after the beeps that indicate the conclusion of the cycle, the contents of the chamber will still be hot. 

Never

  touch  the  internal  parts  of  the  autoclave  directly chamber,  trays,  materials,  etc.

ait  for  them  to  be 

  (

)  when  hot,  w

cool enough before handling. R

 gloves for thermal protection

emember to use safety

.

h

g

T

 

CANCEL

 key for 2 seconds. It’s not recommended to disconnect 

urn the autoclave off after use. To do that, press the

the equipment from the outlet so that it is not necessary to adjust the time and date again.

15

 

For the 

Unwrapped Instruments

 cycle, after the depressurization, the autoclave will show the message:

 

 

 

 

Press the 

Start

 key to proceed with drying. By pressing the 

CANCEL

 key, there will be no drying, 

(flash sterilization). 

Important!

 This option must be used only for immediate use of the instruments.

dry ?

st : yes   Can : no

Note:

 For the autoclave model Vitale Plus 21, when selecting the 

Surgical Kit

 cycle, the surgical case must be placed in 

the central tray. For the Vitale Plus 12, the case (which must be compatible with the size of the autoclave) must be put on 
the top tray.

Summary of Contents for Vitale Plus 12

Page 1: ...uce its environmental impacts and continuous training of its employees achieving this way a sustainable profitability and the maximization of the company s value Rev 2 Develop innovative solutions to...

Page 2: ......

Page 3: ...rist foli s Biosafety Consultant who provides a valuable assistance in the coordination of Crist foli s Biosafety Project complementary products research training of our employees representatives and...

Page 4: ...ymbols Introduction Important Safety Information Installation Instructions Autoclave Components Identification Safety Devices Cycles and Messages Safety Notes How to Use the Autoclave Vitale Plus Alti...

Page 5: ...nding Protection Terminal Fusible Autoclavable Date of manufacture Boas Pr ticas de Fabrica o 4 P B F LOT SN Fragile Handle with care Dangerous Electrical Tension Serial Number Batch code Keep away fr...

Page 6: ...ation we recommend extra attention to this manual Autoclave sterilization although quicker and more efficient requires totally different procedures and handling while still keeping a simple operation...

Page 7: ...y kind of use of this equipment other than the ones described in this manual SPECIAL MEASURES AND OBSERVATIONS DURING THE USE OF THE AUTOCLAVE Always make sure the door handle is properly locked befor...

Page 8: ...he use of an exclusive circuit breaker for the outlet where you intend to install your autoclave If after following all the specifications the electrical current is still oscillating contact your loca...

Page 9: ...of the external steam exit 8 Fig 1A Fig 3 WATER V CUT DO NOT USE THE LID KEEP THE HOSE ABOVE THE WATER LEVEL 5 liters 1 3 gallons Table 2 X NEVER CUT THE CENTRAL PIN OFF Fig 1 EXTERNAL STEAM EXIT VSPF...

Page 10: ...use interference in the cycle EXTERNAL STEAM EXIT Located in the back part of the autoclave Fig 1 page 8 it has a diameter of 5 16 for the connection of the discharge hose which is then connected to t...

Page 11: ...7 Door Gasket Fig 5 Display Start Key Mode Key Cancel Key Lid Panel Keyboard Handle Fig 4 CRIST FOLI A U T O C L A V E S I N S T R U C T I O N M A N U A L Fig 8 Fig 10 Fig 9 250 ml 50 ml 200 ml 150 m...

Page 12: ...onents Identification Figs 6 and 7 page 10 VSPF VALVE Valve that works as an anti vacuum system patent pending FUSE Safety device which purpose is to protect the building electrical wiring against pea...

Page 13: ...an no perform cleaning see manual perform biological test cooling 00 00 T 000 C p 0 00 cooling 00 00 cooling T 000 C power outage subpressure detected cancelled by user overpressure detected temp x pr...

Page 14: ...ymbol 14 appears in some places on the autoclave This means those items require special attention and that the user must observe their references in the Instruction Manual provided with the equipment...

Page 15: ...re for the new cycle is reached the autoclave will resume operation showing heating up When the solenoid valve shuts down the display will show the temperature and pressure rising gradually When the a...

Page 16: ...the CANCEL key to return to the program selection mode ATTENTION Even after the beeps that indicate the conclusion of the cycle the contents of the chamber will still be hot Never touch the internal...

Page 17: ...me ALTITUDE CALIBRATION In order for the autoclave to work properly it is necessary to perform the altitude calibration procedure Calibrate the altitude of the autoclave as shown below for this proced...

Page 18: ...valve closed this will prevent the liquid from boiling over inside the containers when the liquids sterilization cycle was selected If the operator needs to interrupt cancel the heating sterilization...

Page 19: ...ature of drained water Total heat in joules transmitted in one hour Proper working temperature range Proper working altitude 12 liters Aluminum 19 4 kg including components Stainless steel 21 2 kg inc...

Page 20: ...eplace the air filter of the VSPF valve monthly Replace the thermal paste of the heating element annually The component lid Item 2 page 9 must be replaced each 10 years Replace the internal hoses annu...

Page 21: ...hamber Preventive cleaning stainless steel chamber Door gasket cleaning Door gasket replacement Preventive Maintenance of the VSPF valve and replacement of the internal safety valve Air filter replace...

Page 22: ...pose of the label located in the back of the equipment is to identify the autoclave technical data ATTENTION The removal of the identification label and or any other labels stickers affixed to the pro...

Page 23: ...on See How to Proceed when Service is Needed Page 24 THE AUTOCLAVE DOES NOT SWITCH ON POSSIBLE CAUSES SOLUTION The operator connected the autoclave to the power source but didn t press the Start key N...

Page 24: ...g the sterilization of 26 envelopes at a time 10 x 23 cm containing 6 instruments each Remember to leave some space between the envelopes to allow good steam circulation and optimize drying THE AUTOCL...

Page 25: ...ive Maintenance Item 7 Page 19 Crist foli Equipamentos de Biosseguran a Ltda makes clear its concern already demonstrated by the implementation of the Environmental Management System according to stan...

Page 26: ...rcuits fire etc Damage caused by humidity excessive exposure to sun light and salinity Sinister theft or robbery Improper use installation of the equipment and or plugging it into an incorrect voltage...

Reviews: