18
Set Vibrator Activation Point
Controls the point at which the vibrator kicks in within a cycle when the indication while pulling setting is set to
‘Enabled’. There will be 3 vibrations that happen at different times in the cycle. This helps the user understand
where in the cycle they are at a given time. The smaller this figure the earlier in the cycle these vibration points
begin.
Change Trace Length
The trace length is setting the sample rate and the number of samples take within a given time period. The
maximum number of samples take in any cycle will be 1000. To capture the full cycle and obtain the best
resolution it is best to set the trace length as close to the time it takes to complete the cycle. See example
below.
Case 1
– User pulls for 1 second (cycle length), Trace length set to 4 seconds.
Number of samples captured = 250.
Sample Interval = 4ms
Case 2
– User pulls for 1 second (cycle length), Trace length set to 1 seconds. (Optimum)
Number of samples captured = 1000.
Sample Interval = 1ms
Case 3
– User pulls for 4 seconds (cycle length), Trace length set to 1 seconds.
Number of samples captured = 1000. (first second measured only)
Sample Interval = 1ms