INPUT
The flow processor has inputs for turbine meters (both single and dual rotor) and
temperature The flow processor calculates mass flow rates based on volumetric
flow with temperature compensation
Analog Input
You can configure the analog input for different types of temperature transducers
The flow processor is supplied with a 10 kΩ thermistor but can be
jumper-configured for 0…5V DC, 0…10V DC or 4…20 mA temperature
transducers Consult the factory if you need to incorporate the analog input for
reasons other than temperature acquisition See
and
“Wiring and Jumpers” on page 12
to select a particular temperature transducer