 Model RFG2 Service Manual

Summary of Contents for VNUS RFG2

Page 1: ...VNUS Radio Frequency Generator RFGPLUS Model RFG2 Service Manual...

Page 2: ...iance 8 Biomedical Department Power OutputTest 12 Power DeliveryVerification 13 Alarm Priority 13 Error Codes 14 Service and Maintenance 16 Cleaning and Disinfection 16 Repairs and Returns 16 Warranty...

Page 3: ...ator The RF Switch circuitry contained on the UI Board provides the switch and colored backlighting for the RF on off switch located on the right of the FPI as viewed from the front of the Generator R...

Page 4: ...nly be removed with a tool Service Port is UL1950 50VDC Normal MAINSV in Single Fault Condition Metal Enclosure is Protective Earth Display has no touch screen and sits behind protective plastic cover...

Page 5: ...TL100 119 02 1 Set the line voltage to 264V and set the frequency to 50Hz DO NOT EXCEED 280V 2 Plug the line cord of theVNUS Generator into the leakage tester for all tests 3 For source leakage and i...

Page 6: ...tion Nearby electronic equipment may be affected RF emissions CISPR 11 Class A The RFG2 is suitable for use in all establishments other than domestic and those directly connected to the public low vol...

Page 7: ...e considered If the measured field strength in the location in which the RF Generator is used exceeds the applicable RF compliance level above the RF Generator should be observed to verify normal oper...

Page 8: ...nce thermometer and the ESU analyzer Turn off main power to the generator Remove the modified Instrument Cable from the RF Generator and AC power cord from the generator In assessing the functionality...

Page 9: ...es Turn off the unit wait a few seconds and then turn on the unit If the error persists try a different MMC Error 26 The inserted Multimedia Card MMC cannot upgrade this RF Generator because its seria...

Page 10: ...tative will assist the operator with obtaining service or returning the RF Generator as appropriate The RF Generator must be unplugged before servicing WARNING Do not remove the cover of the RF Genera...

Page 11: ...e SJT16 3 IEC color code maximum length 10 ft 3 m Plug minimum 10 A 125VAC Unit receptacle IEC female minimum 10 A 125VAC 220 240VAC Cable H05VVF3G1 0VDE maximum length 10 ft 3 m Plug minimum 6 A 125V...

Page 12: ...0 10 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 40 Peak Voltage V Power W Figure 8 Peak Output Voltage vs Power Output 0 0 5 5 10 10 15 15 20 20 25 30 35 40 25 30 35 40 Power Output W Max Power Se...

Page 13: ...88 797 VEIN 8346 Fax 1 408 365 8470 Covidien Ireland Limited IDA Business Technology Park Tullamore 0123 Customer Service Tel 1 800 962 9888 MEDICAL EQUIPMENT MODEL RFG2 WITH RESPECT TO ELECTRICAL SHO...