Cosman G4 Radiofrequency Generator Service Manual (SPI
11291
Rev A)
Section 0
THERMAL & PULSED RF OUTPUT (continued)
Volts:
0 to 70 volts RMS into 100
Ω
load during Thermal RF output.
0 to 100 volts peak into 100
Ω
during Pulsed RF output.
Resolution: 1 volt RMS during Thermal RF output
1 volt peak during Pulsed RF output
Accuracy:
± 10% of full scale
Current:
0 to 700 mA RMS
Resolution:
1 mA RMS during Thermal RF output
1 mA peak during Pulsed RF output
Accuracy:
± 10% of full scale
PULSED RF SETTINGS:
Repetition Rate: 1, 2, 3, 4, 5, 6, 7, 8 PPS (Hz)
Burst Duration: 0.5, 1, 5, 10, 20, 30 ms
TEMPERATURE MONITOR
Range:
0 – 110°C
Resolution:
1
°C
Accuracy:
± 3 °C between 40 and 90 °C
± 4 °C between 10 and 40 °C, and between 90 and 110 °C
GENERAL
Size: W 14.25” (36.2 cm), H 9.5” (24.1 cm), D 12.15” (30.9 cm)
Weight: 22 lbs (10 kg)
Summary of Contents for G4
Page 2: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 1 ...
Page 66: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 67: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 68: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 69: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 70: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 71: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 72: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 73: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 74: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 75: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 77: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 78: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 80: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 81: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 82: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 85: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 86: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...