Cosman G4 Radiofrequency Generator Service Manual (SPI
11291
Rev A)
Section 7
7.1 Electrode Setup Settings
The Electrode Setup settings determine which electrode output connections (Section 2.1) are to be
delivered Stimulation, Thermal RF, and/or Pulsed RF output.
•
RF Type
: This setting determines the polarity of active electrode output connections.
•
Standard:
(a.k.a. "Monopolar") A Reference Ground Pad must be used for return Currents from
active electrodes flow into Reference Ground Pad Connection (G).
•
Bipolar
: Return Currents flow from active electrode output connections into other active
electrode output connections. Connections 2 and 3 are at reference potential in this
configuration. Voltage, Current, and Impedance readings are displayed for “Bipolar Pairs” of
electrodes, i.e. “Electrode 1+2” and “Electrode 3+4”, of which one is at reference potential and
the other is at potential indicated by the displayed Voltage reading.
•
Active Electrodes
: The number of electrode output connections to which active electrodes should
be attached.
•
Labels:
Text may be assigned to each electrode if RF Type = Standard, and to each bipolar
electrode pair if RF Type = Bipolar. Labels are displayed on the Stimulation, Thermal RF, and
Pulsed RF screens, and are logged to the Summary Record.
The current Electrode Setup settings are indicated using the following labeling in the upper-right corner of
the Stimulation, Thermal RF, Pulsed RF, and Notes screens, as well as in the Summary Record.
RF Type
Active
Electrodes
Electrode Setup Indicator
Standard 1
One
Electrode
Standard
2
Two Electrodes
Standard 3
Three
Electrodes
Standard
4
Four Electrodes
Bipolar
2
Two Electrodes: One Bipolar Pair
Bipolar
4
Four Electrodes: Two Bipolar Pairs
Summary of Contents for G4
Page 2: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 1 ...
Page 66: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 67: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 68: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 69: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 70: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 71: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 72: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 73: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 74: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 75: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 77: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 78: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 80: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 81: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 82: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 85: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 86: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...