Cosman G4 Radiofrequency Generator Service Manual (SPI
11291
Rev A)
8.1.2 Stimulation Settings
The Electrode Setup settings (Section 7.1) determine the configuration of electrodes when the generator
delivers Sensory and Motor Stimulation output. The "Electrode Select Buttons" on the Stimulation screen
(see Section 8.1) are used to select one electrode at a time, or one bipolar electrode pair at a time, from
those set as "active" on the Electrode Setup screen. Stimulation output is delivered only to the one selected
electrode or bipolar electrode pair. The following settings determine the Stimulation output characteristics.
•
Control Units (“Units”)
:
•
Voltage
: Sensory and Motor Stimulation output is Voltage-regulated.
•
Current
: Sensory and Motor Stimulation output is Current-regulated.
•
Auto Ramp
:
•
Off
: The Sensory and Motor Stimulation output level is controlled manually by the user with
the Output Control Knob
(Graphing Interface Only)
.
•
On
: The Sensory and Motor Stimulation output level increases automatically after the user
has turned the output on.
•
Sensory Pulse Rate (“Pulse Rate”):
The rate of Sensory Stimulation output pulses in units of Hertz
(Hz), which is equivalent to units of “Pulses Per Second”.
•
Sensory Pulse Width (“Pulse Width”):
The width of Sensory Stimulation output pulses in units of
milliseconds.
•
Sensory Auto Ramp Max:
The maximum output level if Auto Ramp = On.
•
Motor Pulse Rate (“Pulse Rate”):
The rate of Motor Stimulation output pulses in units of Hertz
(Hz), which is equivalent to units of “Pulses Per Second”.
•
Motor Pulse Width (“Pulse Width”):
The width of Motor Stimulation output pulses in units of
milliseconds.
•
Motor Auto Ramp Max:
The maximum Motor Stimulation output level if Auto Ramp = On.
The Stimulation Settings Screen is shown below.
Select settings from drop-
down menus
Stimulation
Screen
Section 8.1
Custom Setting
Keypad
Select “Custom”
from a setting
drop-down menu
Press “Ok”:
Accept selected
settings.
Section 8
Summary of Contents for G4
Page 2: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 1 ...
Page 66: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 67: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 68: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 69: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 70: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 71: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 72: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 73: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 74: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 75: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 77: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 78: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 80: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 81: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 82: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 85: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 86: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...