Operating Instructions
revision: 06/01
X75 Fusion Splicer
Preparatiss
2-6
-
Close the flap (4) at the cleaver and press it down slowly and carefully. Open the flap
before removing the cleaved fiber in order to avoid damaging the fiber end face.
NOTE
If you press the flap down too quickly, you will obtain a poor quality fiber end face and the
diamond blade may be damaged.
Inserting the Fiber
250
µ
m Coating
-
Open the electrode flap (pos. 1), the holder flap (pos. 2) and the clamping flaps
(pos. 3).
-
Place the fiber (pos. 4) into the v-groove so that the end of the fiber lies between the
electrode tips.
NOTE
Make sure that the fiber end is visible at the monitor after it has been inserted. Poor
splicing quality may otherwise result, or the positioning elements may not demonstrate
adequate travel.
-
First close the holder flap (pos. 2).
-
Hold the fiber straight to ensure that it lies within the groove of the clamping mecha-
nism. Then close the clamping flap (pos. 3).
-
Now insert the second fiber on the other side in the same manner.
-
Close the electrode flap (pos. 1).