10
AXYGEN MICROTUBE SHAKER
5.5 Restore the Factory Settings
1. Turn the instrument On. While the instrument is in self-check mode, press Temp and Prog
buttons at the same time to enter into Calibration mode. NOTE: The instrument cannot
enter into Calibration mode if the self-check is complete.
2. After entering the Calibration mode, press both the Prog and Time buttons to reset, the
interface will display RESET Success, and the instrument will restart automatically.
6.0 Trial Run
Make sure the required operating voltage and power supply voltage match.
Ensure the socket is properly grounded.
Add the medium into the vessel.
Place the vessel on the block.
Plug in the power cord, ensure the power is On and begins initializing.
Secure the heating block for the experiment.
Set the target parameters or programming.
Start working.
Observe the LCD display.
Stop working, and power Off.
If these operations above are normal, the device is ready to operate. If not, the device may be
damaged during transportation, please contact Corning Life Sciences for technical support.
7.0 Troubleshooting
Problem
Solution
Instruments will
not power On
Check whether the power line is unplugged.
Check whether the fuse is broken or loose.
Fault in power On
self-test
Turn Off the unit, then turn On the unit, and reset the instrument to
factory default settings.
If these faults are not resolved, please contact Corning Life Sciences.
8.0 Maintenance and Cleaning
Proper maintenance can keep the instrument working properly and lengthen its lifetime.
Do not
spray cleanser or laboratory alcohols into the instrument when cleaning.
Unplug the power line when cleaning.
Only use recommended cleansers:
–
Dyes: Isopropyl alcohol
–
Construction materials: Water containing tenside/Isopropyl alcohol