AXYGEN MICROTUBE SHAKER
Complete N number of steps Then, press the knob to start work.
Prog shows the program and step number. Refer to the figure above, 2-2 refers to the total
number of steps and the current step of the run.
NOTE:
Before the Time is set, the step will not take effect. If Time is set to 00:00, it will run in
continuous mode. Complete the current step input values prior to programming the next step.
5.3.3 Program with Saving mode
Follow programming practices as mentioned in 5.3.2 for Multiple Step procedures. After the
program has been built, press and hold, program name characters will flash (refer to the red
square in below figure). Rotate the knob to organize each program by a P1-P9 naming system.
Input the preferred P# of the newly generated program and keep for your records, then press
the knob to complete the input. The program saving is complete now.
Prog shows P1, 5-2, means program number 1 has been selected. P1 was saved with 5 steps
and currently working through Step 2.
5.3.4 Load Program
While powered On, press and hold the Prog button. This enters the operator into the Load
program mode. Rotate the knob to show the saved program, from P1 to the last, at most 9
programs. Switch to the program you want, and press the Prog button to check the program
details. Then, press the knob to start or run the program.
5.3.5 Single Function Running
If Temp is set to ---.-, it means working without temperature control function. When the Time
is adjusted to 00:00, the product will run in continuous mode. If speed is set to -----, it means
work without mixing function.