Instruction Manual | 5
Model
Volume
Range
(μL)
Cat. No.
Volume
(μL)
Accuracy
(%)
Precision
(%)
Fit to Tips
Without
Filter
(μL)
Single-channel Pipettors (continued)
5,000
AP-5000
Min. 1,000
±0.6
≤0.25
5,000
2,500
±0.6
≤0.20
Max. 5,000
±0.5
≤0.15
10,000 AP-10ML
Min. 1,000
±2.5
≤0.60
10,000
5,000
±0.8
≤0.30
Max. 10,000
±0.5
≤0.20
8-channel and 12-channel Pipettors
8 - 10
12 - 10
AP-8-10
AP-12-10
Min.
0.5
±10.0
≤8.0
10
5
±4.0
≤2.0
Max.
10
±2.0
≤1.2
8 - 50
12 - 50
AP-8-50
AP-12-50
Min.
5
±4.0
≤2.5
200
25
±3.0
≤1.2
Max.
50
±1.6
≤0.6
8 - 200
12 - 200
AP-8-200
AP-12-200
Min.
20
±3.0
≤1.5
200
100
±1.5
≤0.8
Max.
200
±1.0
≤0.6
8 - 300
12 - 300
AP-8-300
AP-12-300
Min.
50
±1.6
≤1.5
300
150
±1.2
≤1.0
Max.
300
±1.0
≤0.6
The accuracy and precision are obtained gravimetrically using
Axygen® pipet tips, performing at least 10 measurements
of distilled water at a reference temperature of 20°C ± 1°C
according to EN ISO 8655 standard.
2. Setting the Volume
The volume of the pipettor is set by turning the pipetting
pushbutton (A2) or the black adjustment knob (B).
The volume display consists of three numbers and is
read from top to bottom. The three numbers indicate the
volume selected and are colored black and red.
Red numbers represent decimal fractions for microvolume
pipettors or milliliters for macrovolume pipettors.