before Attaching the Adapter

The following steps 

must

be completed prior to attaching the

Adapter to the UPS™ arm:

1. Patient is positioned, prepped and the Manipulator Handle 

is inserted.

2. UPS arm and patient are properly draped.

Attaching the Adapter to the uPS Arm

1. Align the reinforced punch-out hole in the UPS Sterile Drape with

the Mating Hole in the UPS arm.

2. Insert the Adapter’s Engagement Pin (see Figure 1) into the UPS

arm’s Mating Hole until secure. Use the Alignment Pin for proper
orientation of the Adapter.

3. Ensure that the Adapter is securely in place on the arm.

before Attaching the manipulator Handle

The following steps 

must

be completed prior to attaching the

Manipulator Handle to the Adapter:

1. Patient is positioned for surgery.

2. Draped UPS arm and Adapter are placed in approximate position

for attachment near the proximal end of the Manipulator Handle.

Attaching the manipulator Handle

CAuTION: 

DO NOT attach the Manipulator Handle to the UPS arm

until after the patient is in final position and sedated.

1. Open the Latch on the Adapter (see Figure 2).

CAuTION: Securely hold the uPS handle before the foot 
pedal is depressed to avoid inadvertent movement of the
manipulator Handle.

Note:

For optimal motion, grasp only the handle of the UPS to move

the device.

2. Grasp the handle of the UPS, press the foot pedal, and move the

arm to align the Adapter with the Manipulator Handle. Take care
to minimize movement of the Manipulator Handle within the
patient.

3. Position the Mating Pin on the Adapter into the Mating Hole of the

Manipulator Handle.

4. Close and lock the Latch, and ensure that the Manipulator Handle is

securely attached to the Adapter with no free movement.

Removing the manipulator Handle from the Adapter

To detach the Manipulator Handle, unlock and open the Latch on the
Adapter while holding the Manipulator Handle and separate it from 
the Adapter.

Note:

There is no need to depress the foot pedal and move the UPS

arm to separate the Manipulator Handle from the Adapter.

Removing the Adapter from the uPS

To detach the Adapter from the UPS Arm after surgery is complete,
press the Release Button on the Arm (see Figure 3) and pull the
Adapter straight out without rotation.

RECOmmENDED CLEANING INSTRuCTIONS

1. Prepare the enzymatic detergent according to the manufacturers

recommendations and submerge the Adapter in the detergent. 
Soak for ten (10) minutes.

2. Scrub the submerged Adapter with a soft sponge or brush and 

agitate. Clean crevices, using a pipe cleaner/brush or flush with a
syringe when appropriate. Actuate any moving parts to loosen
trapped soil.

3. Rinse in warm (100 °F – 120 °F, 38 °C – 49 °C) tap water for 

one (1) minute. Thoroughly flush difficult to reach areas. 
Actuate while rinsing. 

4. Place the Adapter into a bath containing warm (100 °F – 120 °F, 

38 °C – 49 °C) water. Agitate the Adapter by hand for at least 
one (1) minute. Repeat this process two (2) additional times.

5. Ultrasonically clean the Adapter for twenty (20) minutes in a neutral

pH detergent (Neutrad

®

or acceptable alternative). Prepare the

detergent according to the manufacturer’s recommendations.

6. Rinse the Adapter with clean, tap water, actuating moving parts

while rinsing, for one (1) minute; repeat rinse twice.

7. Dry the Adapter thoroughly with a clean, lint free cloth.
8. Repeat this cleaning procedure if the Adapter appears to be soiled

after initial cleaning and until there is no sign of soil on the device.

RECOmmENDED STEAm AuTOCLAVE STERILIZATION
PARAmETERS

Pre-vacuum Cycle 

Preconditioning Pulses: 4
Exposure temperature: 270 °F – 273 °F  (132 °C – 134 °C)
Exposure time:  4 minutes

Gravity Displacement Cycles

Exposure temperature: 270 °F – 275 °F  (132 °C – 135 °C)
Exposure time:  15 minutes 

Exposure temperature: 250 °F – 253 °F  (121 °C – 123 °C)
Exposure time:  40 minutes

figure 2.  Attaching the Advincula Arch Handle

back View

front View

figure 3.  uPS Arm

DIRECTIONS fOR uSE

figure 1.  Advincula Arch™ Adapter

Mating Pin

Advincula Arch Handle

Advincula

Arch Adapter

Mating Hole

Latch

Mating Pin

Alignment Pin

Engagement Pin

Release Button

Arm

UPS Handle

Latch

CAuTION

: When securing the Latch, use care to avoid contacting

the patient or other accessories in the surgical field.

Summary of Contents for Advincula Arch 53469

Page 1: ...ipulator Handle Refer to the UPS Directions for Use for further instructions ALWAYS position the patient and the OR table prior to attaching the UPS to the Manipulator Handle ALWAYS have the patient u...

Page 2: ...e Manipulator Handle from the Adapter Removing the Adapter from the uPS To detach the Adapter from the UPS Arm after surgery is complete press the Release Button on the Arm see Figure 3 and pull the A...

Page 3: ...e Manipulator Handle from the Adapter Removing the Adapter from the uPS To detach the Adapter from the UPS Arm after surgery is complete press the Release Button on the Arm see Figure 3 and pull the A...

Page 4: ...ipulator Handle Refer to the UPS Directions for Use for further instructions ALWAYS position the patient and the OR table prior to attaching the UPS to the Manipulator Handle ALWAYS have the patient u...

Page 5: ...aut US Bundesgesetz ist der Verkauf dieses Produkts nur an rzte oder im Auftrag von rzten erlaubt Neutrad ist eine eingetragene Marke von Decon Labs Inc CooperSurgical ist eine eingetragene Marke von...

Page 6: ...igen Reinigung immer noch verschmutzt aussieht und bis keine weiteren Anzeichen von Verschmutzung am Instrument vorhanden sind EmPfOHLENE DAmPfAuTOKLAV STERILISATIONSPARAmETER Vorvakuum Zyklus Pr kond...

Page 7: ...igen Reinigung immer noch verschmutzt aussieht und bis keine weiteren Anzeichen von Verschmutzung am Instrument vorhanden sind EmPfOHLENE DAmPfAuTOKLAV STERILISATIONSPARAmETER Vorvakuum Zyklus Pr kond...

Page 8: ...aut US Bundesgesetz ist der Verkauf dieses Produkts nur an rzte oder im Auftrag von rzten erlaubt Neutrad ist eine eingetragene Marke von Decon Labs Inc CooperSurgical ist eine eingetragene Marke von...

Page 9: ...ositivo PRECAuCI N Las leyes federales de EE UU limitan la venta de este dispositivo a m dicos o por prescripci n m dica Neutrad es una marca comercial registrada de Decon Labs Inc CooperSurgical es u...

Page 10: ...c o Pulsos de preacondicionamiento 4 Temperatura de exposici n 132 134 C 270 273 F Tiempo de exposici n 4 minutos Ciclos de desplazamiento por gravedad Temperatura de exposici n 132 135 C 270 275 F Ti...

Page 11: ...c o Pulsos de preacondicionamiento 4 Temperatura de exposici n 132 134 C 270 273 F Tiempo de exposici n 4 minutos Ciclos de desplazamiento por gravedad Temperatura de exposici n 132 135 C 270 275 F Ti...

Page 12: ...ositivo PRECAuCI N Las leyes federales de EE UU limitan la venta de este dispositivo a m dicos o por prescripci n m dica Neutrad es una marca comercial registrada de Decon Labs Inc CooperSurgical es u...

Page 13: ...ne marque d pos e de Decon Labs Inc CooperSurgical est une marque d pos e de CooperSurgical Inc Toutes les autres marques avec les symboles TM et sont des marques de commerce et des marques d pos es d...

Page 14: ...PEuR Cycle vide pr alable Impulsions de pr conditionnement 4 Temp rature d exposition 132 134 C 270 273 F Dur e d exposition 4 minutes Cycles d placement de gravit Temp rature d exposition 132 135 C 2...

Page 15: ...PEuR Cycle vide pr alable Impulsions de pr conditionnement 4 Temp rature d exposition 132 134 C 270 273 F Dur e d exposition 4 minutes Cycles d placement de gravit Temp rature d exposition 132 135 C 2...

Page 16: ...ne marque d pos e de Decon Labs Inc CooperSurgical est une marque d pos e de CooperSurgical Inc Toutes les autres marques avec les symboles TM et sont des marques de commerce et des marques d pos es d...

Page 17: ...ENZIONE La legge federale statunitense limita la vendita di questo dispositivo ai medici o su presentazione di prescrizione medica Neutrad un marchio depositato di Decon Labs Inc CooperSurgical un mar...

Page 18: ...a gravit Temperatura di esposizione 132 C 135 C 270 275 F Tempo di esposizione 15 minuti Temperatura di esposizione 121 C 123 C 250 253 F Tempo di esposizione 40 minuti figura 2 Attacco dell impugnatu...

Page 19: ...a gravit Temperatura di esposizione 132 C 135 C 270 275 F Tempo di esposizione 15 minuti Temperatura di esposizione 121 C 123 C 250 253 F Tempo di esposizione 40 minuti figura 2 Attacco dell impugnatu...

Page 20: ...ENZIONE La legge federale statunitense limita la vendita di questo dispositivo ai medici o su presentazione di prescrizione medica Neutrad un marchio depositato di Decon Labs Inc CooperSurgical un mar...

Page 21: ...DO A lei federal dos EUA s permite a venda deste dispositivo sob receita m dica Neutrad uma marca registada da Decon Labs Inc CooperSurgical uma marca registada da CooperSurgical Inc Todas as outras m...

Page 22: ...v cuo Impulsos de pr condicionamento 4 Temperatura de exposi o 132 C 134 C 270 F 273 F Tempo de exposi o 4 minutos Ciclos de deslocamento de gravidade Temperatura de exposi o 132 C 135 C 270 F 275 F T...

Page 23: ...v cuo Impulsos de pr condicionamento 4 Temperatura de exposi o 132 C 134 C 270 F 273 F Tempo de exposi o 4 minutos Ciclos de deslocamento de gravidade Temperatura de exposi o 132 C 135 C 270 F 275 F T...

Page 24: ...DO A lei federal dos EUA s permite a venda deste dispositivo sob receita m dica Neutrad uma marca registada da Decon Labs Inc CooperSurgical uma marca registada da CooperSurgical Inc Todas as outras m...

Reviews: