background image

13  

I-ZDEG-EU-1312-394-03

ENGLISH

zENITH® TX2® DISSECTIoN ENDovaSCULar 
GrafT WITH Pro-form® aND THE z-Trak® PLUS 
INTroDUCTIoN SySTEm

read all instructions carefully. failure to properly follow the instructions, 
warnings, and precautions may lead to serious consequences or injury to 
the patient.
CaUTIoN: federal (U.S.a.) law restricts this device to sale by or on the order 
of a physician.
CaUTIoN: all contents of the inner pouch (including the introduction 
system and endovascular graft) are supplied sterile, for single use only.

1 DEVICE DESCRIPTION
1.1 Zenith TX2 Dissection Endovascular Graft with Pro-Form and 
the Z-Trak Plus Introduction System

When treating dissections of the descending thoracic aorta, the Zenith TX2 
Dissection Endovascular Graft with Pro-Form and the Z-Trak Plus Introduction 
System is typically used in conjunction with the Zenith® Dissection Endovascular 
Stent. The Zenith TX2 Dissection Endovascular Graft with Pro-Form and the 
Z-Trak Plus Introduction System is for sealing of entry tear(s). The Zenith 
Dissecton Endovascular Stent provides support to delaminated segments of 
the aorta. For information regarding the use and deployment of the Zenith 
Dissection Endovascular Stents, please refer to the Instructions for Use.
The Zenith TX2 Dissection Endovascular Graft with Pro-Form and the Z-Trak 
Plus Introduction System shortest Straight Components can be used to provide 
additional length to the endovascular graft, both proximal and distal. Ensure 
there is a minimum overlap of 2 stents. The stent graft is constructed of full-
thickness woven polyester fabric sewn to self-expanding stainless steel Cook-Z 
stents with braided polyester and monofilament polypropylene suture (fig. 1). 
The Zenith TX2 Dissection Endovascular Graft with Pro-Form and the Z-Trak Plus 
Introduction System is fully stented to provide stability and the expansile force 
necessary to open the lumen of the graft during deployment. Additionally, the 
Cook-Z stents provide the necessary attachment and seal of the graft to the 
vessel wall.
To facilitate fluoroscopic visualization of the stent graft, four radiopaque 
markers are positioned on each end of the Straight Component or Tapered 
Component. These markers are placed in a circumferential orientation within 1 
mm of the most proximal aspect of the graft material and within 1 mm of the 
most distal aspect of the graft material.

1.2 Thoracic Z-Trak Plus Introduction System

The Zenith TX2 Dissection Endovascular Graft with Pro-Form and the Z-Trak Plus 
Introduction System is shipped preloaded onto the Z-Trak Plus Introduction 
System. It has a sequential deployment method with built-in features to pro-
vide continuous control of the endovascular graft throughout the deployment 
procedure. The Z-Trak Plus Introduction System enables precise positioning 
before deployment of the Straight Component or Tapered Component. The 
straight and tapered graft components are deployed from a 20 French or 22 
French Z-Trak Plus Intro duc tion System. These systems use a single trigger-
wire release mechanism to secure the endovascular graft onto the delivery 
introduction system until released by the physician (fig. 2). All introduction 
systems are compatible with a .035 inch wire guide. For added hemostasis, the 
Captor® Hemostatic Valve can be loosened or tightened for the introduction 
and/or removal of ancillary devices into and out of the sheath. All introduction 
systems feature Flexor® introducer sheaths, which resist kinking and are 
hydrophilically coated. Both features are intended to enhance trackability in the 
iliac arteries and thoracic aorta.
To facilitate sheath withdrawal, each graft component is kept in a longitudinally 
stretched condition on the introduction system by locking trigger-wires (fig. 3). 
These trigger-wires work in tandem to deliver sequential controlled release of 
the Zenith TX2 Dissection Endovascular Graft with Pro-Form during deployment 
(fig. 3).

2 INTENDED USE

The Zenith TX2 Dissection Endovascular Graft with Pro-Form and the Z-Trak Plus 
Introduction System is indicated for the endovascular treatment of patients 
with symptomatic dissection of the descending thoracic aorta having vascular 
morphology suitable for endovascular repair (fig. 4), including:

•  Adequate iliac/femoral access compatible with the required introduction 

systems,

•  Radius of curvature greater than 35 mm along the length of aorta intended 

to be treated by Straight or Tapered Component,

•  Non-dissected/aneurysmal aortic segments (fixation sites) proximal to the 

entry tear:
•  with a length of at least 20 mm,
•  with a diameter measured outer wall to outer wall of no greater than 38 

mm and no less than 20 mm, and

•  with localized angulation less than 45 degrees.

3 CONTRAINDICATIONS

The Zenith TX2 Dissection Endovascular Graft with Pro-Form and with the Z-Trak 
Plus Introduction System is contraindicated in:

•  Patients with known sensitivities or allergies to stainless steel, polyester, 

polypropylene, nitinol or gold.

•  Patients with a systemic infection who may be at increased risk of 

endovascular graft infection.

4 WARNINGS AND PRECAUTIONS
4.1 General

•  Read all instructions carefully. Failure to properly follow the instructions, 

warnings, and precautions may lead to serious consequences or injury to the 
patient.

•  The Zenith TX2 Dissection Endo vascular Graft with Pro-Form and the Z-Trak 

Plus Introduction System should only be used by physicians and teams 
trained in vascular interventional techniques (catheter-based and surgical) 
and in the use of this device. Specific training expectations are described in 
Section 10.1, Physician Training.

•  The long-term performance of endovascular grafts has not yet been 

established. All patients should be advised that endovascular treatment 
requires life-long, regular follow-up to assess their health and the 
performance of their endovascular graft. Patients with specific clinical 
findings (e.g., endoleaks, enlarging aneurysms, persisting flow in false lumen, 
or changes in the structure or position of the endovascular graft) should 
receive enhanced follow-up. Specific follow-up guidelines are described in 
Section 12, ImaGING GUIDELINES aND PoSToPEraTIvE foLLoW-UP.

•  After endovascular graft placement, patients should be regularly monitored 

for perigraft flow or changes in the structure or position of the endovascular 
graft. At a minimum, annual imaging is required, including: 1) thoracic 
device radiographs to examine device integrity (separation between 
components or stent fracture); and 2) contrast and non-contrast CT to 
examine perigraft flow, patency, tortuosity, device position and progressive 
disease. If renal complications or other factors preclude the use of image 
contrast media, use of other imaging modalities (e.g., TEE, IVUS) should be 
considered in combination with non-contrast CT.

•  The Zenith TX2 Dissection Endo vascular Graft with Pro-Form and the 

Z-Trak Plus Introduction System is not recommended in patients unable to 
undergo, or who will not be compliant with, the necessary preoperative and 
postoperative imaging and implantation studies as described in Section 12, 
ImaGING GUIDELINES aND PoSToPEraTIvE foLLoW-UP.

•  Additional endovascular interventions or conversion to standard open 

surgical repair following initial endovascular repair should be considered for 
patients experiencing unacceptable decrease in fixation length (vessel and 
component overlap) and/or endoleak.

•  Patients experiencing reduced blood flow through the graft and/or leaks 

may be required to undergo secondary endovascular interventions or 
surgical procedures.

•  Always have a qualified surgery team available during implantation or 

reintervention procedures in the event that conversion to open surgical 
repair is necessary.

•  Interventions such as defibrillation, cardioversion, or CPR, although not 

specifically evaluated in studies, may have the potential to disrupt position 
or seal of the endograft, and should be followed by imaging to confirm 
continued device function.

4.2 Patient Selection, Treatment and Follow-Up

•  Access vessel diameter (measured inner wall to inner wall) and morphology 

 (tortuosity, oc clusive disease, and/or calcification) should be compatible 
with vascular access techniques and introduction systems of the profile 
of a 20 French or 22 French vascular introducer sheath. Vessels that are 
significantly calcified, occlusive, tortuous or thrombus-lined may preclude 
femoral introduction of the endovascular graft and/or may increase the risk 
of embolization.

•  Key anatomic elements that may affect successful exclusion of the dissection 

include severe angulation (localized angulation > 45 degrees); short 
proximal fixation site (< 20 mm); an inverted funnel shape at the proximal 
fixation site (greater than 10% increase in diameter over 20 mm of fixation 
site length); and circumferential thrombus and/or calcification at the arterial 
fixation sites. Irregular calcification and/or plaque may compromise the 
attachment and sealing at the fixation site. Necks exhibiting these key 
anatomic elements may be more conducive to graft migration.

•  The Zenith TX2 Dissection Endo vascular Graft with Pro-Form and the Z-Trak 

Plus Introduction System is not recommended for patients who cannot 
tolerate contrast agents necessary for intraoperative and postoperative 
follow-up imaging.

•  The Zenith TX2 Dissection Endo vascular Graft with Pro-Form and the Z-Trak 

Plus Introduction System is not recommended for patients whose weight or 
size would compromise or prevent the necessary imaging requirements.

•  Graft implantation may increase the risk of paraplegia where graft exclusion 

covers the origins of dominant spinal cord or intercostal arteries.

•  Patients with connective tissue disorders have not been evaluated.
•  Highly patent intercostal aortic branches or large collateral vessels are 

likely to result in retrograde flow after thoracic graft implantation. Patients 
with uncorrectable coagulopathy may also have an increased risk of Type II 
endoleak or bleeding complications.

4.3 Implant Procedure

•  Systemic anticoagulation should be used during the implantation 

procedure based on hospital and physician preferred protocol. If heparin is 
contraindicated, an alternative anticoagulant should be used.

•  Minimize handling of the constrained endoprosthesis during preparation 

and insertion to decrease the risk of endoprosthesis contamination and 
infection.

•  To activate the hydrophilic coating on the outside of the sheath, the surface 

must be wiped with sterile gauze pads soaked in saline solution. Always 
keep the sheath hydrated for optimal performance.

•  Maintain wire guide position during introduction system insertion.
•  Do not bend or kink the introduction system. Doing so may cause damage 

to the introduction system and the Zenith TX2 Dissection Endovascular Graft 
with Pro-Form.

•  Always use fluoroscopy for guidance, delivery, and observation of the 

Zenith TX2 Dissection Endovascular Graft with Pro-Form and the Z-Trak Plus 
Introduction System within the vasculature.

•  The use of the Zenith TX2 Dissection Endovascular Graft with Pro-Form and 

the Z-Trak Plus Introduction System requires administration of intravascular 
contrast. Patients with pre-existing renal insufficiency may have an increased 
risk of renal failure postoperatively. Care should be taken to limit the amount 
of contrast media used during the procedure.

•  To avoid twisting the endovascular graft, never rotate the introduction 

system during the procedure. Allow the device to conform naturally to the 
curves and tortuosity of the vessels.

•  As the sheath is withdrawn, anatomy and graft position may change. 

Constantly monitor graft position and perform angiography to check 
position as necessary.

•  Inaccurate placement and/or incomplete sealing of the Zenith TX2 

Dissection Endovascular Graft with Pro-Form within the vessel may result 
in increased risk of endoleak, migration, or inadvertent occlusion of the left 
subclavian, left common carotid, and/or celiac arteries.

•  Incorrect deployment or migration of the endoprosthesis may require 

surgical intervention.

•  Do not continue advancing the wire guide or any portion of the introduction 

system if resistance is felt. Stop and assess the cause of resistance; vessel, 
catheter, or graft damage may occur. Exercise particular care in areas of 
stenosis, intravascular thrombosis, or calcified or tortuous vessels.

•  Unless medically indicated, do not deploy the Zenith TX2 Dissection 

Endovascular Graft with Pro-Form in a location that will occlude arteries 
necessary to supply blood flow to organs or extremities. Do not cover 
significant arch or mesenteric arteries (exception may be the left subclavian 
artery) with the endoprosthesis. Vessel occlusion may occur. If a left 
subclavian artery is to be covered with the device, the clinician should 
be aware of the possibility of compromise to cerebral and upper limb 
circulation.

•  Use caution during manipulation of catheters, wires and sheaths within a 

dissection. Significant disturbances may dislodge fragments of thrombus, 
which can cause distal or cerebral embolization.

•  Avoid damaging the graft or disturbing graft positioning after placement 

in the event reinstrumentation (secondary intervention) of the graft is 
necessary.

•  Do not attempt to re-sheath the graft after partial or complete deployment.
•  Repositioning the stent graft distally after partial deployment of the covered 

proximal stent may result in damage to the stent graft and/or vessel injury.

•  To avoid impaling any catheters left in situ, rotate the introduction system 

during withdrawal.

4.4  Molding Balloon Use - Optional

•  Confirm complete deflation of balloon prior to repositioning.
•  Do not inflate balloon in aorta outside of graft.
•  Use caution during molding within a dissection
•  For added hemostasis, the Captor™ Hemostatic Valve can be loosened or 

tightened to accomodate the insertion and subsequent withdrawal of a 
molding balloon.

Summary of Contents for Zenith TX2

Page 1: ...Zenith TX2 Dissection Endovascular Graft with Pro Form and the Z Trak Plus Introduction System Instructions for Use EN 13 394 03 I ZDEG EU 1312 394 03 Patient I D Card Included ...

Page 2: ...ith 15 10 2 Inspection Prior to Use 15 10 3 Materials Required 15 10 4 Device Diameter Sizing Guidelines 15 Table 1 Straight Component and Tapered Component Graft Diameter Sizing Guide 15 11 DIRECTIONS FOR USE 15 General Use Information 15 Pre Implant Determinants 15 Patient Preparation 15 11 1 The Zenith TX2 Dissection Endovascular Graft with Pro Form and the Z Trak Plus Intro duction System Prep...

Page 3: ...lustrações Illustrationer 1 3 EN a Zenith TX2 Dissection Endovascular Graft with Pro Form and the Z Trak Plus Introduction System b Radiopaque markers c Tapered component d Straight component EN a Pre Tensioned Zenith TX2 Dissection Endovascular Graft with Pro Form a a b 2 EN a Thoracic Z Trak Plus Introduction System b Straight component tapered component introduction system a b c d ...

Page 4: ...trações Illustrationer 4 EN a Zenith TX2 Dissection Endovascular Graft with Pro Form and Zenith Dissection Endovascular Stent b proximal neck length 20 mm c proximal neck diameter 20 38 mm d Zenith Dissection Endovascular Graft e aortic radius 35 mm f Zenith Dissection Endovascular Stent g distal aortic diameter 20 38 mm h aortic radius 35 mm endovascular stent component a b c d e f g h ...

Page 5: ... EU 1312 394 03 Illustrations Ilustrace Illustrationer Abbildungen Απεικονίσεις Ilustraciones Illustrations Illusztrációk Illustrazioni Afbeeldingen Illustrasjoner Ilustracje Ilustrações Illustrationer 5 7 9 8 6 ...

Page 6: ...ar Graft with Pro Form and the Z Trak Plus Introduction System is not recommended in patients unable to undergo or who will not be compliant with the necessary preoperative and postoperative imaging and implantation studies as described in Section 12 IMAGING GUIDELINES AND POSTOPERATIVE FOLLOW UP Additional endovascular interventions or conversion to standard open surgical repair following initial...

Page 7: ...ive procedure than open surgical repair Therefore potential clinical benefits to patients treated with the Zenith TX2 Dissection Endovascular Graft with Pro Form may include a suitable dissection repair with less risk and fewer complications than those treated with open surgical repair Zenith TX2 Dissection Endovascular Graft with Pro Form patients may benefit from a reduced risk of serious treatm...

Page 8: ...raft Diameter3 mm Overall Length of Straight Component mm Overall Length of 4 mm Tapered Component mm Overall Length of 8 mm Tapered Component mm Introducer Sheath Fr 20 22 79 117 20 21 24 79 117 20 22 23 26 79 136 20 24 28 82 142 202 20 25 30 82 142 202 20 26 30 82 142 202 20 27 30 82 142 202 20 28 32 82 142 202 162 202 158 196 20 29 32 82 142 202 162 202 158 196 20 30 34 79 154 204 159 199 156 1...

Page 9: ...m that there are no endoleaks or kinks and verify position of proximal and distal gold radiopaque markers Remove the sheaths wires and catheters NOTE If endoleaks or other problems are observed refer to Section 11 2 Additional Devices 3 Repair vessels and close in standard surgical fashion 11 2 Additional Devices Inaccuracies in device size selection or placement changes or anomalies in patient an...

Page 10: ...3 0 Tesla Temperature Rise In nonclinical testing the Zenith TX2 TAA Endovascular Graft produced a temperature rise of less than or equal to 1 3 C scaled to an SAR of 2 0 W kg during 15 minutes of MR imaging i e for one scanning sequence performed in a MR 3 0 Tesla System General Electric Excite HDx Software G3 0 052B Image Artifact The image artifact extends throughout the anatomical region conta...

Page 11: ...hopni nebo ochotni podstoupit nezbytné předoperační a pooperační zobrazovací vyšetření a implantační studie viz popis v části 12 POKYNY PRO SNÍMKOVÁNÍ A POOPERAČNÍ KONTROLU U pacientů u nichž se vyskytne neakceptovatelné zkrácení délky fixace překrytí cévy s komponentou a nebo endoleak je třeba zvážit další endovaskulární intervenci nebo přechod na standardní otevřenou chirurgickou operaci následn...

Page 12: ... pravděpodobně méně invazivní zákrok než otevřená chirurgická operace Proto lze mezi potenciální klinické přínosy u pacientů léčených pomocí endovaskulárního graftu Zenith TX2 Dissection s prvkem Pro Form zahrnout vhodnou reparaci disekce s menším rizikem a s méně komplikacemi ve srovnání s otevřenou chirurgickou operací Pacienti s endovaskulárním graftem Zenith TX2 Dissection s prvkem Pro Form mo...

Page 13: ... mm Zaváděcí sheath French 20 22 79 117 20 21 24 79 117 20 22 23 26 79 136 20 24 28 82 142 202 20 25 30 82 142 202 20 26 30 82 142 202 20 27 30 82 142 202 20 28 32 82 142 202 162 202 158 196 20 29 32 82 142 202 162 202 158 196 20 30 34 79 154 204 159 199 156 194 20 31 36 79 154 204 159 199 159 199 22 32 36 79 154 204 159 199 159 199 22 33 38 79 154 204 154 204 159 199 22 34 38 79 154 204 154 204 1...

Page 14: ...my viz část 11 2 Dodatečná zařízení 3 Sešijte cévy a uzavřete pole standardními chirurgickými technikami 11 2 Dodatečná zařízení Nepřesnost při výběru velikosti zařízení nebo při jeho umístění změny nebo anomálie anatomických poměrů pacienta nebo komplikace při operaci mohou vyžadovat umístění dalších endovaskulárních graftů Bez ohledu na to které zařízení bylo umístěno základní výkon bude podobný...

Page 15: ... snímkovací sekvenci v MRI systému o 1 5 tesla Siemens Magnetom software Numaris 4 Zvýšení teploty při 3 0 tesla V neklinických testech došlo u endovaskulárního graftu Zenith TX2 TAA ke zvýšení teploty o max 1 3 C přepočítané na SAR 2 0 W kg za 15 minut snímkování MRI tj za jednu snímkovací sekvenci v MRI systému o 3 0 tesla General Electric Excite HDx software G3 0 052B Artefakt obrazu Při snímko...

Page 16: ...Manufacturer WILLIAM COOK EUROPE ApS Sandet 6 DK 4632 Bjaeverskov DENMARK www cookmedical com COOK 2013 ...

Reviews: