background image

16 

I-ZDEG-EU-1312-394-03

•  Appropriate size sheath (e.g., 5.0 French)
•  Pigtail flush catheter (often radiopaque-banded sizing catheters; i.e., Cook 

Centimeter Sizing CSC-20 catheter)

2.  Perform angiography at the appropriate level. If using radiopaque markers, 

adjust position as necessary and repeat angiography.

3.  Ensure graft system has been flushed and primed with heparinized saline 

(appropriate flush solution), and all air has been removed.

4.  Give systemic heparin. Flush all catheters and wet all wire guides with a 

strong heparin solution. This should be repeated following each exchange.

5.  Replace the standard wire guide with a stiff .035 inch, 260/300 cm LESDC 

wire guide and advance through the catheter and up to the aortic arch.

6.  Remove pigtail flush catheter and sheath.
  NoTE: At this stage, the second femoral artery can be accessed for 

angiographic catheter placement. Alternatively, a brachial approach may be 
considered.

7.  Introduce the freshly hydrated introduction system over the wire guide and 

advance until the desired graft position is reached.

  CaUTIoN: To avoid twisting the endovascular graft, never rotate the 

introduction system during the procedure. allow the device to conform 
naturally to the curves and tortuosity of the vessels.

 NoTE: 

The dilator tip will soften at body temperature.

 NoTE: 

To facilitate introduction of the wire guide into the introduction 

system, it may be necessary to slightly straighten the introduction system 
dilator tip.

8.  Verify wire guide position in the aortic arch. Ensure correct graft position.
9.  Ensure that the Captor Hemostatic Valve on the Flexor Introducer Sheath is 

turned to the open position (fig. 8).

10.  Stabilize the gray positioner (introduction system shaft) and withdraw the 

sheath until the graft is fully expanded and the valve assembly docks with 
the control handle.

 

CaUTIoN: As the sheath is withdrawn, anatomy and graft position may 
change. Constantly monitor graft position and perform angiography to 
check position as necessary.

 NoTE: If extreme difficulty is encountered when attempting to withdraw 

the sheath, place the device in a less tortuous position that enables the 
sheath to be retracted. Very carefully withdraw the sheath until it just 
begins to retract, and stop instantly. Move back to original position and 
continue deployment.

11.  Verify graft position and adjust it forward, if necessary. Recheck graft 

position with angiography.

 

 NoTE: If an angiographic catheter is placed parallel to the stent graft, use 
this to perform position angiography.

12.  Loosen the safety lock from the green trigger-wire release mechanism. 

Withdraw the trigger-wire in a continuous movement until the proximal 
end of the graft opens (fig. 9). Do not rotate the green trigger-wire knob. 
Withdraw the trigger-wire completely to release the distal attachment to 
the introducer.

 NoTE: Check to make sure that all trigger-wires are removed prior to 

withdrawal of the introduction system.

13.  Remove the introduction system, leaving the wire guide in the graft.
 NoTE: Leave the Zenith TX2 Dissection Endovascular Graft with Pro-Form 

and the Z-Trak Plus Introduction System in place if intending to use a 
dissection stent.

11.1.2 Molding Balloon Insertion - Optional

1.  Prepare molding balloon as follows and/or per the manufacturer’s 

instructions.

•  Flush wire lumen with heparinized saline.
•  Remove all air from balloon.

2.  In preparation for the insertion of the molding balloon, open the Captor 

Hemostatic Valve by turning it counter-clockwise.

3.  Advance the molding balloon over the wire guide and through the 

hemostatic valve of the main body introduction system to the level of the 
proximal fixation site. Maintain proper sheath positioning.

4.  Tighten the Captor Hemostatic Valve around the molding balloon with 

gentle pressure by turning it clockwise.

5.  Expand the molding balloon with diluted contrast media (as directed by the 

manufacturer) in the area of the proximal covered stent, starting proximally 
and working in the distal direction.

  CaUTIoN: Do not inflate balloon in aorta outside of graft. Use caution 

during molding within a dissection.

  CaUTIoN: Confirm complete deflation of balloon prior to 

repositioning.

6.  Open the Captor Hemostatic Valve, remove the molding balloon 

and replace it with an angiographic catheter to perform completion 
angiograms.

7.  Tighten the Captor Hemostatic Valve around the angiographic catheter 

with gentle pressure by turning it clockwise.

8.  Remove or replace all stiff wire guides to allow aorta to resume its natural 

position.

Final Angiogram

1.  Position angiographic catheter just above the level of the endovascular 

graft. Perform angiography to verify correct positioning. Verify patency of 
arch vessels and celiac plexus.

2.  Confirm that there are no endoleaks or kinks, and verify position of 

proximal and distal gold radiopaque markers. Remove the sheaths, wires 
and catheters.

  NoTE: If endoleaks or other problems are observed, refer to Section 11.2, 

additional Devices.

3.  Repair vessels and close in standard surgical fashion.

11.2 Additional Devices

Inaccuracies in device size selection or placement, changes or anomalies 
in patient anatomy, or procedural complications can require placement of 
additional endovascular grafts. Regardless of the device placed, the basic 
procedure(s) will be similar to the maneuvers required and described previously 
in this document. It is vital to maintain wire guide access.

12 IMAGING GUIDELINES AND POSTOPERATIVE FOLLOW-UP
12.1 General

The long-term performance of endovascular grafts has not yet been 
established.
 All patients should be advised that endovascular treatment 
requires life-long, regular follow-up to assess their health and performance of 
their endovascular graft and/or stent. Patients with specific clinical findings (e.g., 
endoleaks, persisting flow in false lumen or changes in the structure or position 
of the endovascular graft) should receive additional follow-up. Patients should 
be counseled on the importance of adhering to the follow-up schedule, both 
during the first year and at yearly intervals thereafter. Patients should be told 
that regular and consistent follow-up is a critical part of ensuring the ongoing 
safety and effectiveness of endovascular treatment of dissections.
Physicians should evaluate patients on an individual basis and prescribe their 
follow-up relative to the needs and circumstances of each individual patient. 
The recommended imaging schedule is presented in Table 2. This schedule 
continues to be the minimum requirement for patient follow-up and should 
be maintained even in the absence of clinical symptoms (e.g., pain, numbness, 
weakness). Patients with specific clinical findings (e.g., endoleaks, enlarging 
aneurysms, or changes in the structure or position of the stent graft or stent) 
should receive follow-up at more frequent intervals.
Annual imaging follow-up should include thoracic device radiographs and both 
contrast and non-contrast CT examinations. If renal complications or other 
factors preclude the use of image contrast media, thoracic device radiographs 
and non-contrast CT may be used.

•  The combination of contrast and non-contrast CT imaging provides 

information on device migration, endoleak, patency, tortuosity, progressive 
disease, fixation length, and other morphological changes.

•  The thoracic device radiographs provide information on device integrity 

(separation between components and stent fracture).

Table 2 lists the minimum requirements for imaging follow-up for patients with 
the Zenith TX2 Dissection Endovascular Graft with Pro-Form and the Z-Trak 
Plus Introduction System. Patients requiring enhanced follow-up should have 
interim evaluations.

Table 2 Recommended Imaging Schedule for Endograft Patients

angiogram

CT

(contrast and non-contrast)

Thoracic Device radiographs

Pre-procedure

X

1

Procedural

X

Pre-discharge (within 7 days)

X

2

X

1 month

X

2

X

6 month

X

2

X

12 month (annually thereafter)

X

2

X

1

 Imaging should be performed within 6 months before the procedure.

2

 If Type I or III endoleak, prompt intervention and additional follow-up post-intervention recommended, See

 Section 12.5, additional Surveillance and Treatment.

Table 3 Acceptable Imaging Protocols

Non-contrast

Contrast

IV contrast

No

Yes

Acceptable machines

Spiral capable of > 40 seconds

Spiral capable of > 40 seconds

Injection volume

n/a

150 cc

Injection rate

n/a

> 2.5 cc/sec

Injection mode

n/a

Power

Bolus timing

n/a

Test bolus: Smart Prep, C.A.R.E. or equivalent

Coverage - start

Neck

Subclavian aorta

Coverage - finish

Diaphragm

Profunda femoris origin

Collimation

< 3 mm

< 3 mm 

Reconstruction

2.5 mm throughout - soft algorithm

2.5 mm throughout - soft algorithm

Axial DFOV

32 cm

32 cm

Post-injection runs

None

None

12.2 Contrast and Non-Contrast CT Recommendations

•  Film sets should include all sequential images at lowest possible slice 

thickness (≤ 3 mm). Do NOT perform large slice thickness (> 3 mm) and/or 
omit consecutive CT images/film sets, as it prevents precise anatomical and 
device comparisons over time.

•  All images should include a scale for each film/image. Images should be 

arranged no smaller than 20:1 images on 14” x 17” sheets if film is used.

•  Both non-contrast and contrast runs are required, with matching or 

corresponding table positions.

•  Pre-contrast and contrast run slice thickness and interval must match.
•  DO NOT change patient orientation or re-landmark patient between non-

contrast and contrast runs.

Non-contrast and contrast enhanced baseline and follow-up imaging are 
important for optimal patient surveillance. It is important to follow acceptable 
imaging protocols during the CT exam. Table 3 lists examples of acceptable 
imaging protocols.

Summary of Contents for Zenith TX2

Page 1: ...Zenith TX2 Dissection Endovascular Graft with Pro Form and the Z Trak Plus Introduction System Instructions for Use EN 13 394 03 I ZDEG EU 1312 394 03 Patient I D Card Included ...

Page 2: ...ith 15 10 2 Inspection Prior to Use 15 10 3 Materials Required 15 10 4 Device Diameter Sizing Guidelines 15 Table 1 Straight Component and Tapered Component Graft Diameter Sizing Guide 15 11 DIRECTIONS FOR USE 15 General Use Information 15 Pre Implant Determinants 15 Patient Preparation 15 11 1 The Zenith TX2 Dissection Endovascular Graft with Pro Form and the Z Trak Plus Intro duction System Prep...

Page 3: ...lustrações Illustrationer 1 3 EN a Zenith TX2 Dissection Endovascular Graft with Pro Form and the Z Trak Plus Introduction System b Radiopaque markers c Tapered component d Straight component EN a Pre Tensioned Zenith TX2 Dissection Endovascular Graft with Pro Form a a b 2 EN a Thoracic Z Trak Plus Introduction System b Straight component tapered component introduction system a b c d ...

Page 4: ...trações Illustrationer 4 EN a Zenith TX2 Dissection Endovascular Graft with Pro Form and Zenith Dissection Endovascular Stent b proximal neck length 20 mm c proximal neck diameter 20 38 mm d Zenith Dissection Endovascular Graft e aortic radius 35 mm f Zenith Dissection Endovascular Stent g distal aortic diameter 20 38 mm h aortic radius 35 mm endovascular stent component a b c d e f g h ...

Page 5: ... EU 1312 394 03 Illustrations Ilustrace Illustrationer Abbildungen Απεικονίσεις Ilustraciones Illustrations Illusztrációk Illustrazioni Afbeeldingen Illustrasjoner Ilustracje Ilustrações Illustrationer 5 7 9 8 6 ...

Page 6: ...ar Graft with Pro Form and the Z Trak Plus Introduction System is not recommended in patients unable to undergo or who will not be compliant with the necessary preoperative and postoperative imaging and implantation studies as described in Section 12 IMAGING GUIDELINES AND POSTOPERATIVE FOLLOW UP Additional endovascular interventions or conversion to standard open surgical repair following initial...

Page 7: ...ive procedure than open surgical repair Therefore potential clinical benefits to patients treated with the Zenith TX2 Dissection Endovascular Graft with Pro Form may include a suitable dissection repair with less risk and fewer complications than those treated with open surgical repair Zenith TX2 Dissection Endovascular Graft with Pro Form patients may benefit from a reduced risk of serious treatm...

Page 8: ...raft Diameter3 mm Overall Length of Straight Component mm Overall Length of 4 mm Tapered Component mm Overall Length of 8 mm Tapered Component mm Introducer Sheath Fr 20 22 79 117 20 21 24 79 117 20 22 23 26 79 136 20 24 28 82 142 202 20 25 30 82 142 202 20 26 30 82 142 202 20 27 30 82 142 202 20 28 32 82 142 202 162 202 158 196 20 29 32 82 142 202 162 202 158 196 20 30 34 79 154 204 159 199 156 1...

Page 9: ...m that there are no endoleaks or kinks and verify position of proximal and distal gold radiopaque markers Remove the sheaths wires and catheters NOTE If endoleaks or other problems are observed refer to Section 11 2 Additional Devices 3 Repair vessels and close in standard surgical fashion 11 2 Additional Devices Inaccuracies in device size selection or placement changes or anomalies in patient an...

Page 10: ...3 0 Tesla Temperature Rise In nonclinical testing the Zenith TX2 TAA Endovascular Graft produced a temperature rise of less than or equal to 1 3 C scaled to an SAR of 2 0 W kg during 15 minutes of MR imaging i e for one scanning sequence performed in a MR 3 0 Tesla System General Electric Excite HDx Software G3 0 052B Image Artifact The image artifact extends throughout the anatomical region conta...

Page 11: ...hopni nebo ochotni podstoupit nezbytné předoperační a pooperační zobrazovací vyšetření a implantační studie viz popis v části 12 POKYNY PRO SNÍMKOVÁNÍ A POOPERAČNÍ KONTROLU U pacientů u nichž se vyskytne neakceptovatelné zkrácení délky fixace překrytí cévy s komponentou a nebo endoleak je třeba zvážit další endovaskulární intervenci nebo přechod na standardní otevřenou chirurgickou operaci následn...

Page 12: ... pravděpodobně méně invazivní zákrok než otevřená chirurgická operace Proto lze mezi potenciální klinické přínosy u pacientů léčených pomocí endovaskulárního graftu Zenith TX2 Dissection s prvkem Pro Form zahrnout vhodnou reparaci disekce s menším rizikem a s méně komplikacemi ve srovnání s otevřenou chirurgickou operací Pacienti s endovaskulárním graftem Zenith TX2 Dissection s prvkem Pro Form mo...

Page 13: ... mm Zaváděcí sheath French 20 22 79 117 20 21 24 79 117 20 22 23 26 79 136 20 24 28 82 142 202 20 25 30 82 142 202 20 26 30 82 142 202 20 27 30 82 142 202 20 28 32 82 142 202 162 202 158 196 20 29 32 82 142 202 162 202 158 196 20 30 34 79 154 204 159 199 156 194 20 31 36 79 154 204 159 199 159 199 22 32 36 79 154 204 159 199 159 199 22 33 38 79 154 204 154 204 159 199 22 34 38 79 154 204 154 204 1...

Page 14: ...my viz část 11 2 Dodatečná zařízení 3 Sešijte cévy a uzavřete pole standardními chirurgickými technikami 11 2 Dodatečná zařízení Nepřesnost při výběru velikosti zařízení nebo při jeho umístění změny nebo anomálie anatomických poměrů pacienta nebo komplikace při operaci mohou vyžadovat umístění dalších endovaskulárních graftů Bez ohledu na to které zařízení bylo umístěno základní výkon bude podobný...

Page 15: ... snímkovací sekvenci v MRI systému o 1 5 tesla Siemens Magnetom software Numaris 4 Zvýšení teploty při 3 0 tesla V neklinických testech došlo u endovaskulárního graftu Zenith TX2 TAA ke zvýšení teploty o max 1 3 C přepočítané na SAR 2 0 W kg za 15 minut snímkování MRI tj za jednu snímkovací sekvenci v MRI systému o 3 0 tesla General Electric Excite HDx software G3 0 052B Artefakt obrazu Při snímko...

Page 16: ...Manufacturer WILLIAM COOK EUROPE ApS Sandet 6 DK 4632 Bjaeverskov DENMARK www cookmedical com COOK 2013 ...

Reviews: