MAINTENANCE & INSPECTION
• The operator of the lift shall inspect the ConvaLift before each use. Check
all bolts for tightness. Make sure the base can be easily widened, and
that all lift parts are in place. Make sure that casters can be turned
freely, and that caster brakes can be engaged. Make certain all
necessary items (i.e. slings, and wheelchairs, etc.) are ready for use.
• At least once a month, the lift should be thoroughly inspected by a
person qualified to recognize any signs of wear and tear, and looseness
of bolts or parts. Replace any worn parts immediately.
• To lubricate, put a drop of oil on the following points when the ConvaLift is
placed into service and every month thereafter: Top of Mast, Spreader
Hinge, and Caster Axles.
FIRST
RECEIVED
MONTHLY
EVERY 3
MONTHS
BOOM & SPREADER BAR
Check connections between Boom and
Spreader Bar for improper connection,
looseness, or wear
Check connections between Boom and Mast for
improper connection, looseness, or wear.
Check the Boom for bending and deflection.
MAST
Check the Mast for bending or deflection.
BASE & FOOT PEDAL
Check bolts and nuts for looseness..
Check casters and axle bolts for tightness.
Check rubber parts on the casters for deflection