Control Components Cerlic MultiTracker Manual Download Page 8

 

 

 

 

 

   Multi 

Tracker 

 

 

 

 

 

 

 

 

Description of the Display in Profile Mode 

When pressing the Profile button marked 

 the display will show the sludge profile. 

See fig 3 below. At any time during a measurement it is possible to switch between text 
mode and profile mode by pressing the button 

 without affecting the measuring 

value. General information such as sludge concentration and depth is always shown in 

either the text or profile modes windows.   

 
 

 

 

 

 

 

 

 

 

 

 

 

 

                                                                                       

 

fig 3 

 
 

 

 

 
 
 

 

 

 
 

 

 

 
 

 

 

 

 fig 4 

 

 

Profile displayed in x=mg/l and 

  y=depth 

 

Solids Concentration 

 

 

Depth – current 

 

 

Fluff level

 

 

Blanket depth

 

 Maximum depth 

 

 

Sensor position during depth 

  measuring versus max. depth

 

 
 
 
 
 

Summary of Contents for Cerlic MultiTracker

Page 1: ...2018 02 14 C84F5EN19 Multi Tracker MultiTracker MultiTracker s n xxx xx xxx Blanko s n xxx xx xxx Solido s n xxx xx xxx OxyDuo s n xxx xx xxx DO cell s n xxx xx xxx...

Page 2: ...NCENTRATION UNITS 11 Depth units 11 LOG 11 6 GETTING STARTED 12 SAVE A MEASUREMENT 12 DOWNLOADING DATA FROM THE MULTITRACKER CONTROLLER 13 SOFTWARE UPDATE 13 7 MAINTENANCE 14 INSPECTION OF THE SENSOR...

Page 3: ...N MENU 30 Settings 30 Calibration 32 Getting Started 33 Maintenance 33 SPECIFICATION OXYDUO 34 OXYGEN SOLUBILITY CHART 35 APPENDIX 3 SOLIDO SUSPENDED SOLIDS SENSOR 36 Description of the function 36 Me...

Page 4: ...ate when it comes to recycling and help to save our environment If such waste not is handled and recycled according to regulation EC Directive 2002 96 EC the environment as well as people s health may...

Page 5: ...ssible to wind up all of the cable on the cable hook The cable is held in place on the hook with an O ring see fig 3 fig 2 fig 3 Removing the cable hook To detach the cable hook unsnap the SS sensor f...

Page 6: ...nterface The following icons show the keypad push buttons and describe their functions Opens the main menu or confirms enters a choice in the menus Closes a menu choice without changing or exit s one...

Page 7: ...n like time date and temperature will always be displayed in the top left side of the screen Different sensors attached to the hand unit will display different information on the display The screen sh...

Page 8: ...switch between text mode and profile mode by pressing the button without affecting the measuring value General information such as sludge concentration and depth is always shown in either the text or...

Page 9: ...maximum depth The settings are senor dependent and the screen shoot below is done with a Blanko sensor attached the hand controller fig 9 Preferences In the Preferences menu fig 10 it is possible to...

Page 10: ...nd confirm with Step up with or down with until desired language is found and confirm with Exit the menu with until one of the two main windows appear text mode or profile mode Time and date To set ti...

Page 11: ...Depth units Following depth units are available cm dm m in ft with inches Press to open the menu and step down with Select SETUP with Step down with Select Depth unit and confirm with Step upwith or d...

Page 12: ...nd not in the sensor 6 Getting Started Save a measurement To save a measurement or to perform a new measurement press When doing this an option to save or restart the measurement will be presented The...

Page 13: ...Talk to interface with the hand unit and move the stored data to the PC The software can be downloaded free of charge from the Cerlic home page fig 15 To transfer the stored data disconnect the sensor...

Page 14: ...oft cloth and water Pay attention not to scratch the quartz measuring windows or damage the built in pressure sensor Changing of Batteries Cerlic Blanket Tracker is supplied with four AA batteries pla...

Page 15: ...com The WEB page also has the correct shipping address 9 Spare Parts The MultiTracker is supplied with four AA batteries The batteries are the only parts that are subject to be replaced by the user It...

Page 16: ...an French Log Function 250 Measurements Signal at Preset Alarm Values Acoustic Vibration Display Temperature Range Liquid 0 50 C 32 122 F Temperature Range of hand controller 10 50 C 14 122 F Key Pad...

Page 17: ...ay measured values as text or as a graphic image of the sludge profile Two different alarm levels for the sludge blanket may be set to indicate fluff and sludge blanket levels These may be displayed n...

Page 18: ...the Display in Text Mode At start up the display is shown in fig 1 and description in fig 2 Value for solids concentration and depth are displayed on the left and alarm values for fluff 6 and sludge b...

Page 19: ...o switch between text mode and profile mode by pressing the button without affecting the measuring value General information such as sludge concentration and depth is always shown in either the text o...

Page 20: ...to sludge or solids on top of liquid Meas Mode Depth shows depth from top of liquid to bottom of tank as sensor is lowered Height Starts out at Max Depth and depth values go down as sensor is lowered...

Page 21: ...blanket level In a clarifier this concentration is normally choosen to correspond with the return sludge RAS concentration e g 5 000 mg l ppm and in a thickener e g 7 000 mg l fig 8 Press to open the...

Page 22: ...with or down with in order to select desired mode Confirm with Exit the menu with until one of the two main windos appear text mode or profile mode Setup See setup in section for MultiTracker hand co...

Page 23: ...xposed to light and confirm with When zero calibration is done exit with until one of the two main windows appears text mode or profile mode Conc sample Sludge Sample Press to open the menu and step d...

Page 24: ...er OFF and ON after you have completed zero 1 m depth calibrations This resets the MultiTracker to accept the new depth values Getting Started Getting started with the Blanko Start the MultiTracker by...

Page 25: ...ith or down with to reach desired log position The postion tag will be shown at the bottom of the display window Fig 11 below shows that the actual profile will be stored at position General Fig 11 It...

Page 26: ...er to verify if a depth calibration is needed lower the sensor into water until the one meter cable mark The depth value should not differ more than 2cm 1 A higher deviation requires a zero and one me...

Page 27: ...g l mg l ppm Depth units cm dm m in ft with inches Principe of Depth Measurement Pressure cell absolute pressure Accuracy Depth Measurement 0 5 FS Altitude limitation Max 4 500 feet optional 15 000 f...

Page 28: ...ents with information about time date location that may be named by the user maximum 10 characters e g Clarifier 2 Thickener 5 etc Each location will contain data oxygen level and temperature that app...

Page 29: ...e for oxygen concentration are displayed on the left and alarm values to the right Date time and temperature are shown across the top fig 1 fig 2 Min measured oxygen level Max measured oxygen level Da...

Page 30: ...are outside this Min limit the hand controller will make one acoustic beep signal When oxygen levels are outside this Max limit the hand controller will make two acoustic beep signals Threshold 1 Low...

Page 31: ...reached then this will be shown on the display as High fig 6 Press to open the menu and select SETTINGS confirm with Step down with and select Threshold 2 with Step up with or down with until desired...

Page 32: ...Sample and confirm with Wait for a stable value and confirm with Leave the menu with until the main menu is reached Disconnect the bracket Put the bracket on one pin for future use Replace the protect...

Page 33: ...mmand from the text mode display where it is possible to select position for storing the measurement Step up with or down with to reach desired log position The postion tag will be shown in the lower...

Page 34: ...ic representative In case of sending the Oxyduo to Cerlic for checkup or repair please make sure to use the form for Return of Material RMA prior to dispatch The RMA document can be downloaded from th...

Page 35: ...35 Multi Tracker Oxygen Solubility chart...

Page 36: ...els Measurement Functions It is able to store up to 250 measurements with information about time date location that may be named by the user maximum 10 characters e g Clarifier 2 Thickener 5 etc Each...

Page 37: ...concentration is displayed on the left and alarm values to the right Date time and temperature are shown across the top fig 1 fig 2 Min measured suspended solids conc Max measured suspended solids con...

Page 38: ...ation levels are outside this Min limit then the hand controller will make one acoustic beep signal When suspended solids concentration levels are outside this Max limit the hand controller will make...

Page 39: ...arm value is reached then this will be shown on the display as Max fig 6 Press to open the menu and select SETTINGS confirm with Step down with and select Thrreshold 2 with Step up with or down with u...

Page 40: ...pears Conc sample Sludge Sample Press to open the menu and step down with Select Calibration with Step down with and select Sample Confirm with Put the senor in a sample with known concentration and s...

Page 41: ...t the menu between each recalibration select the next calibration curve and enter the calibration menu again Only one calibration point can be entered for each curve Note The zero water calibration va...

Page 42: ...eavy wires to withstand mechanical wear for a long life The sensor and enclosure cable fittings are high quality MatchClamp to ensure a water proof connection even should the outer shield be damaged I...

Page 43: ...s 1 FS full scale Repeatability 2 of Measured Value Type of Measurement Continuous Number of Detectable Levels Two Measuring units g l mg l ppm Temperature Range Liquid 0 50 C 32 122 F Sensor Body Sta...