41
Multi
Tracker
Three Calibration Curve Setup*
When using the Solido sensor, the Mulitracker supports up to three different calibration
curves. This could be used to improve accuracy when measuring in different types of
sludge´s.
The symbol indicated above shows the selected calibration curve (curve 1 is currently
active). To change curves then you need to be on start-up screen which shows solids
concentration. To change the selected calibration curve, press and hold
while using
the
/
buttons to change curves. Three curves are available.
To modify or recalibrate one of these curves, then first select the calibration curve to be
recalibrated as described above, which must be done before entering the main menu
screen by pressing
and then calibration option.
To recalibrate multiple calibration curves, you will have to exit the menu between each
recalibration, select the next calibration curve, and enter the calibration menu again.
Only one calibration point can be entered for each curve.
Note
The zero (water) calibration value is common to all three calibration curves, and
therefore does only need to be performed for one of them.
*This function is supported with software version MT2.3 and above.
Getting Started
Getting started with the Solido
Start the MultiTracker by pressing the button marked
. To switch OFF/ON the unit,
press the same button. The unit will start automatically. Values for suspended solids are
displayed in clear text and real time. If the unit is not active during an eight minute
period then it will automatically be turn off without saving the measurements.
Save a measurement
To save a measurement, or to perform a new measurement, press
, when doing this,
an option to save
or restart the measurement
will be presented. There is a
short command from the text mode display where it is possible to select position for
General