Multi
Tracker
0 m Cal. - Zero depth calibration
Press
to open the menu and step down with
. Select ”Calibration”
with
. Step down with
and select ”0 m cal. ”. Hold sensor vertically in
the air and start zero calibration with
. Continue until value is stable and
then save calibration with
. Exit the menu with
.
1 m Cal. – 1 m or 39.4” depth calibration
Press
to open the menu and step down with
. Select ”Calibration”
with
. Step down with
and to ”1 m cal. ”. Lower the sensor into the
water until the 1 meter SS collar is level with the top of the liquid. Start
Calibration with
and save calibration when value is stable by once again
pressing
. When all settings are done, exit with
until one of the two
main windows appears (text mode or profile mode).
Must turn Tracker OFF/ON after zero & 1 m depth calibrations
Turn tracker OFF
and ON
after you have completed zero & 1 m
depth calibrations. This resets the MultiTracker to accept the new depth
values.
Getting Started
Getting started with the Blanko
Start the MultiTracker by pressing the button marked
. To switch OFF/ON the unit,
press the same button. The unit will do a atmospheric pressure compensation at start-
up and measurement will start automatically. Values for sludge concentration and
depth are displayed in clear text and real time. If the unit is not active during an eight
minute period then it will automatically be turn off without saving the measurements.
In order to get correct scaling of the depth range in profile range shown on the display,
then it is necessary to set the maximal depth at the actual measuring position.
Setting of Maximum depth
Press
to open the menu and select ”Settings” with
. Step down with
until
desired Max Depth is selected with
. Select the depth value numbers by stepping
and change the values with
or
. Save by pressing
to save.