Problem
Possible Cause Solution
I pressed the
power button and
the device doesn't
turn on
The batteries are
inserted improperly
Replace batteries in proper position
as shown on page 19 of this manual
The batteries are dead
Replace all 4 AAA batteries.
Never mix old and new batteries.
Display is on but
there is no signal
The device is paused
Press ON/Off button
The electode module is
not inserted correctly
Slide the electrode module onto
groves of the M1 device until you
hear it click into place.
The lead wire snaps
may not be connected
to the buttons on the
gel pads
Turn the device off and snap the
lead wire onto the button of the
gel pad
The battery
symbol is flashing
The batteries are low
Replace all 4 AAA batteries.
Never mix old and new batteries.
The contractions
are weak even
though the
insensity is high
The batteries are low
Replace all 4 AAA batteries.
Never mix old and new batteries.
The gel pads are not
making proper contact
Replace gel pads
The pads are positioned
improperly
Turn the M1 device off, remove
the Corebelt™ and repostion as
described on page 20 of this
manual. Make sure the gel pads
are below your rib cage.
Stimulation is
uncomfortable
The gel pads may
be worn out
Replace gel pads
Gel pads may not
be properly
Turn the M1 device off, remove
the Corebelt™ and repostion as
described on page 20 of this
manual. Make sure the gel pads
are below your rib cage.
Too many sessions in
a row
Leave at least six hours
between sessions
Tr
ou
ble
sho
ot
in
g