13
Average value
Calculate the Average value in every 4 measure value.
The deviation between average value and true value does
not exceed 1%.
Pulse
Pulse
Pulse
Pulse Parameter
Parameter
Parameter
Parameter Specification
Specification
Specification
Specification
Measuring range
30bpm
~
250bpm, (the resolution is 1bpm)
Accuracy
±
2bpm or
±
2% (select the larger)
Average pulse rate
Moving calculate the Average pulse rate every 4
cardio-beat
,
s cycle.
The deviation between average value and true value does
not exceed 1%
Safety
Safety
Safety
Safety Type
Type
Type
Type
Interior Battery
,
BF Type
Pulse
Pulse
Pulse
Pulse Intensity
Intensity
Intensity
Intensity
Range
Continuous bar-graph display, the higher display indicate
the stronger pulse.
Battery
Battery
Battery
Battery Requirement
Requirement
Requirement
Requirement
Voltage 3.7 rechargeable lithium battery × 1 (The red wire on the battery denotes anode,the black wire
on the battery denotes cathode.)
Battery
Battery
Battery
Battery working
working
working
working life
life
life
life
Charge and discharge no less than 500 times.
Power
Power
Power
Power Adapter
Adapter
Adapter
Adapter
Input Voltage
100 to 240 VAC, 50/60 Hz
Output voltage
5 VDC
Output current
250mA
Output power
1.25 W
O
O
O
Oximeter
ximeter
ximeter
ximeter P
P
P
Probe
robe
robe
robe
Wavelength:660nm 880nm
Dimensions
Dimensions
Dimensions
Dimensions and
and
and
and Weight
Weight
Weight
Weight
Dimensions
57(L) × 32(W) × 30 (H) mm
Weight
About 50g (with the lithium battery*1)