About Calibration
Although this product is calibrated before shipping, you can use the calibration program to calibrate
analog input and output yourself.
Starting the calibration program
Click the [Calibration] button on the property page for the device to start the calibration program.
Proceed with connecting the calibration equipment and performing the calibration in accordance with
the instructions displayed by the calibration program.
Analog input calibration
Analog input calibration requires a reference voltage generator.
As the analog input has 12-bit resolution, use a reference voltage generator with a precision of at least 4
digits after the decimal point.
Calibrate one channel only for each range that you use.
*This product can calibrate only gain.
Analog output calibration
Analog output calibration requires a digital multimeter.
As the analog output has 12-bit resolution, use a multimeter with a precision of at least 4 digits after the
decimal point.
Calibrate each channel separately for each range that you use.
*This product can calibrate only gain.
Factory setting
You can use the calibration program to restore the factory calibration settings.
If for some reason you are unable to achieve the rated accuracy, please contact your retailer.
* The name of the board
you have just added is
displayed.
- AIO-121602LN-USB
- AIO-120802LN-USB
Summary of Contents for AIO-120802LN-USB
Page 7: ...vi AIO 121602LN USB AIO 120802LN USB ...
Page 17: ...1 Before Using the Product 10 AIO 121602LN USB AIO 120802LN USB ...
Page 53: ...3 Connection and Installation 46 AIO 121602LN USB AIO 120802LN USB ...
Page 63: ...5 About C LOGGER 56 AIO 121602LN USB AIO 120802LN USB ...
Page 93: ...6 Function 86 AIO 121602LN USB AIO 120802LN USB ...
Page 104: ......