2
1.2.2
To avoid fire hazard or electrical
shock:
1.
Prior to each use, the light source and all
associated equipment must be inspected for
proper operation.
2.
Use only non-conducting light guides
(optical cable). Verify that the light guide is
dry and free from cleaning solution residue
before plugging it into the light source.
3.
Use only the power cord specified for this
device.
4.
Do not use plug adapters or extension
cords; such devices defeat the safety
ground.
5.
Do not excessively bend or kink the power
cord. Always inspect cords for signs of
excessive wear or damage. If wear or
damage are found, discontinue use and
replace immediately.
6.
Unplug the power cord from the outlet
when performing any kind of cleaning or
general maintenance, such as changing the
lamp assembly.
7.
Use only the fuse type and rating specified
for this device.
8.
Do not apply an input voltage that is outside
the specified range.
9.
Do not operate this product with the cover
removed. The lamp access door must be
closed and latched for the light source to
work.
10. Do not expose the light source to moisture,
or operate the light source in either damp or
wet areas, or place liquids on or above the
light source.
11. Do not operate the light source in an
explosive atmosphere.
12. Do not allow foreign objects inside the light
source.
13. Do not disassemble the light source. There
are no user-serviceable parts inside.
14. Never spill liquids of any kind on the device
or immerse it in liquids.
15. Only plug the light source into a properly
grounded hospital-grade outlet.
1.2.3
To avoid personal injury and
damage to the light source.
1.
Do not use the light source with
incompatible equipment or accessories that
are not authorized by ConMed Linvatec.
Doing so may void certifications and/or
warranties.
2.
Do not place light source on an unstable
surface, cart, stand or table.
3.
Do not allow the light source to share an
electrical outlet or grounding with life
supporting or life sustaining equipment.
4.
Only connect IEC 60950 or IEC 60601
series compliant devices to signal input or
output ports.
5.
Do not look directly or indirectly at the
lamp when it is on. The lamp produces high
intensity visible light that may cause burns
to skin or eyes.
6.
Do not touch the lamp until it has cooled to
room temperature. Severe burn may result.