13
3.5
Troubleshooting
Symptom
Possible Cause
Corrective Action
Lamp will not ignite.
♦
Power cord is not plugged in or
is faulty.
♦
Power is not turned on.
♦
Lamp access door open.
♦
Lamp module is not fully
engaged.
♦
Lamp is expired.
♦
System overheats or lamp has
expired.
♦
Plug power cord into receptacle
and/or Light Source.
♦
Replace power cord.
♦
Place POWER button in the on
position. LED will illuminate
green when unit is on.
♦
The Light Source will not work
with the lamp access door open.
Assure door is completely closed
and latched.
♦
Verify lamp assembly is
completely engaged on
receptacle pins within Light
Source.
♦
Replace lamp (see
“3.3 Lamp
Replacement” on page 10
).
♦
Turn power off. Ensure intake
vents are not blocked. Allow
system to cool for five minutes.
Turn power back on. If lamp
does not ignite, replace lamp
assembly.
♦
Use of unauthorized lamp
module. Use only ConMed
authorized replacement lamp
modules.