background image

 

11

 

1.

Depress the POWER button to the 
off position.

2.

Unplug the light source. Allow to cool 
before proceeding.

3.

Flip the latch on the lamp access door 
towards the back of the unit and pull the 
door open.

4.

Remove the lamp module by pulling 
outward.

5.

Insert the new lamp module into place until 
it seats.

6.

After installation is complete, close and 
secure the door.

 

NOTE: Follow local governing ordinances 
and recycling plan regarding disposal of 
device components.

Summary of Contents for LS7700

Page 1: ...ConMed Linvatec 2330 Millrace Court Unit 5 Mississauga ON L5N 1W2 For technical support loaner requests please call 1 800 387 9636...

Page 2: ...TABLE OF CONTENTS LS7700 SERVICE DISCLAIMER 3 LS7700 BLOCK DIAGRAM 4 BILL OF MATERIALS 5 LS7700 FINAL CHECKOUT PROCEDURES FCO 7 LS7700 USER MANUAL 11...

Page 3: ...optimum product performance and any deviations may adversely impact the function of the device The risks associated with improper servicing and or calibration may include but are not limited to patien...

Page 4: ...s propietary information and shall not be copied reproduced disclosed to others or used in any way without the prior written consent of ConMed Linvatec ConMed Linvatec Endoscopy 7416 Hollister Avenue...

Page 5: ...390770 1 SUB ASM FRONT PNL LS7700 EA 1 22 5505 SCRW 6 32x1 4 SHCS ALLOY EA 11 23 390770 3 SUB ASM SHUTTER MOTOR EA 1 24 5502 NUT 6 32 KEP SS EA 7 25 18509 PWR SUPPLY 300W LS7700 EA 1 26 90387 FM BRACK...

Page 6: ...EEVE IM4000 EA 4 58 60110 LABEL UL CANADA EA 1 59 61034 LBL SN CHASSIS LS7700 EA 1 60 812300 DOC PROC FCO LS7700 EA 0 61 5035 SCR 4 40X1 4IN 100CSK FL PHH SS EA 3 62 5516 SCRW 6 32 X 1 1 8 PAN SS EA 4...

Page 7: ...0 00 6 0 0 60 5 0 8 3 0 5 1 4 5 2 00 6 0 9 5 4 0 6 4 8 7 0 4 5 00 1 6 1 4 0 4 0 00 4 0 4 5 4 6 00 5 0 5 0 3 C o n f i d e n t i a l 0 60 0 60 5 0 5 0 5 5 1 1 00 6 0 9 5 00 6 0 9 5 4 0 6 4 8 7 4 0 6 4...

Page 8: ...6 8 0 0 0 5 0 0 0 3 6 0 00 4 A B 0 60 0 5 0 0 3 5 5 0 0 5 0 0 00 0 54 0 0 0 0 0 0 0 0 0 0 3 5 A C D 4 6 4 6 0 E 0 0 0 0 4 6 C o n f i d e n t i a l 0 00 4 0 00 4 0 60 5 0 5 0 5 5 0 0 Controlled Docum...

Page 9: ...0 6 5 0 0 0 0 5 4 4 0 4 E 0 4 0 0 0 1 4 4 0 3 5 0 0 0 0 60 A 0 0 D 8 0 0 4 6 G 4 6 4 0 0 0 4 4 6 A 4 0 3 5 C o n f i d e n t i a l 0 0 0 0 A A 0 0 D 8 D 8 0 0 0 0 4 6 G 4 6 G 4 6 4 6 4 4 Controlled D...

Page 10: ...4 00 4 0 0 5 0 3 00 0 0 00 7 5 0 2 3 0 5 3 0 0 0 0 5 6 1 C o n f i d e n t i a l 0 0 0 5 6 0 5 6 Controlled Document Effective 09 Jun 2009...

Page 11: ...The ConMed Linvatec LS7700 300 Watt Xenon Light Source Instruction Manual...

Page 12: ...val system or translated into any human or computer language in any form or by any means electronic mechanical magnetic manual or otherwise or disclosed to third parties in whole or in part without th...

Page 13: ...the wall outlet if any of the following occur 3 1 3 Receiving Inspection 3 1 4 Symbol Definitions 4 1 5 System Features 5 1 5 1 Front Panel 5 1 5 2 Rear Panel 6 2 0 SYSTEM INSTALLATION and OPERATION 2...

Page 14: ...ii Table of Contents Page 4 0 TECHNICAL INFORMATION 4 1 Light Source 16 4 2 Lamp Module 16 4 3 Accessories Compatible Equipment 23 4 4 Customer Service 24 4 4 1 Assistance 24 4 4 2 Repairs 24...

Page 15: ...ean the unit and accessories See 3 2 Cleaning on page 10 1 2 1 Warnings 1 Do not attempt to open or service the light source chassis There are no user serviceable parts inside Removing the cover may i...

Page 16: ...xpose the light source to moisture or operate the light source in either damp or wet areas or place liquids on or above the light source 11 Do not operate the light source in an explosive atmosphere 1...

Page 17: ...device from the wall outlet if any of the following occur 1 The power cord is damaged 2 Liquid has been spilled on the device or it has been exposed to other moisture 3 The device has been dropped or...

Page 18: ...sk of explosion if used in the presence of flammable anesthetics Type BF equipment No user service recommended Refer servicing to qualified ConMed Linvatec service personnel HOT CAUTION Protective Ear...

Page 19: ...e Organic Light Emitting Diode OLED Display Displays the settings of the light source such as stand by mode brightness level lamp life hours remaining and power on or off BRIGHTNESS Buttons Allows the...

Page 20: ...accept a cable from a ConMed Nurse s Assistant NA or Surgeon s Assistant SA to enable external control by an NA or SA CAMERA Connector This connector accepts an ethernet cable only from an optional C...

Page 21: ...removed 2 2 Operation of Light Source Use the following sequence to operate the light source 1 Press the POWER button to turn the light source on The power button LED will illuminate green and the OLE...

Page 22: ...in the OFF condition the POWER button LED illuminates orange Power will continue to run to the unit but the lamp is not on STAND BY and BRIGHTNESS buttons are inactive Pressing the POWER button will...

Page 23: ...ttons are controlled by the IM4000 series camera to achieve the desired picture brightness The STAND BY button returns the light source to STAND BY mode The POWER button will turn the light source OFF...

Page 24: ...ilized water Dry with a clean lint free towel 3 3 Lamp Replacement The following maintenance procedures should only be performed by qualified service personnel Read all warnings and cautionary notices...

Page 25: ...the lamp access door towards the back of the unit and pull the door open 4 Remove the lamp module by pulling outward 5 Insert the new lamp module into place until it seats 6 After installation is comp...

Page 26: ...driver or similar tool gently release the retention tab on the fuse drawer and slide it out 3 Pull out existing fuses and replace only with factory supplied fuses as stated on the rear of the unit Rep...

Page 27: ...utton in the on position LED will illuminate green when unit is on The Light Source will not work with the lamp access door open Assure door is completely closed and latched Verify lamp assembly is co...

Page 28: ...guide fibers are broken Increase lamp brightness Replace light guide Check lamp life indicated on OLED Replace lamp Check to ensure light guide is fully inserted into LGR Inspect light guide cables I...

Page 29: ...e in this current regulated mode until the power is removed by the front panel POWER button The temperature on the surface of the lamp can be very high therefore cooling is very important during opera...

Page 30: ...0 lbs 7 26 kg Dimensions 14 0 W x 5 2 H x 15 1 D 35 5cm W x 13 2cm H x 38 4cm D Power Cable Length 6 ft 2m ENVIRONMENTAL Operating Ambient Temperature 42 F to 86 F 6 C to 30 C Relative Humidity 30 to...

Page 31: ...MA NEMKO OVE SEMKO VDE UTE FEMKO Receptacle End Molded straight PVC plug DIN 49457 CEE 22 V 10A 250V VDE D N S SEV OVE KEMA Cord PVC 7 2mm diameter 10A 250V Conductors 3 x 1 mm2 Conductor Colors brown...

Page 32: ...CISPR 11 Group 1 The LS7700 Light Source uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipme...

Page 33: ...ine to line 2 kV lines to earth 1 kV line to line 2 kV lines to earth Mains power quality should be that of a typical commercial or hospital environment Power frequency 50 60 Hz magnetic field IEC 610...

Page 34: ...um output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as de...

Page 35: ...ite survey should be considered If the measured field strength in the location in which the LS7700 Light Source is used exceeds the applicable RF compliance level above the LS7700 Light Source should...

Page 36: ...ccording to Frequency of Transmitter m Read Maximum Output Power of Transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz W d 1 2 P d 1 2 P d 2 4 P 0 01 0 012 0 012 0 024 0 1 0 12 0 12 0...

Page 37: ...CMI Scope Fitting 7452 Weck Scope Fitting 7462 Storz Olympus Scope Fitting 7453 Wolf Dyonics Light Source Fitting 7455 ACMI Stryker Light Source Fitting 7456 Storz Linvatec Light Source Fitting 7457 O...

Page 38: ...s R G number prominently displayed on the box and included on all paperwork Refer to this number if making inquiries about repair status Please call ConMed Linvatec Customer Service and provide the fo...

Page 39: ...25...

Page 40: ...26...

Page 41: ......

Page 42: ...All rights reserved Printed in USA W41 129 004 Rev AA 06 2007 Linvatec Europe B1070 Brussels Belgium EC REP...

Reviews: