E-46
E
a Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the
VP47XX
and
VP47XXF
Monitors are used exceeds the
applicable RF compliance level above, the
VP47XX
and
VP47XXF
Monitors should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the
VP47XX
and
VP47XXF
Monitors.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The
ConMed Linvatec
VP47XX
and
VP47XXF
Monitors are intended for use in the
electromagnetic environment specified below. The customer or the user of this
VP47XX
/
VP47XXF
Monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environment
guidance
Summary of Contents for Linvatec VP4726
Page 54: ...E 50 E ...
Page 58: ...F iv Table des matières Page F ...
Page 114: ...F 56 F ...
Page 118: ...G iv Inhaltsverzeichnis Seite G ...
Page 170: ...G 52 G ...
Page 174: ...I iv Sommario Pagina I ...
Page 228: ...I 54 I ...
Page 232: ...S iv Contenido Página S ...
Page 285: ......