Minimum recommended sterilization exposure times are as follows:
Table 1: Sterilization Parameters
Sterilization Type
Temperature
Minimum
Exposure
Time
Dry Cycle
Sterilization Guidelines for Sterilizing PowerPro Attachments Individually or in the
PowerPro PRO5090 System Sterilization Case
Steam Pre-vacuum
270 - 272°F (132 - 133°C)
4 minutes
8 minutes minimum *
Steam Gravity
270 - 272°F (132 - 133°C)
10 minutes
8 minutes minimum *
Steam Gravity
250 - 254°F (121 - 123°C)
40 minutes
8 minutes minimum *
Sterilization Guidelines for Sterilizing PowerPro Attachments in the PowerPro Basic
Sterilization Case (Lid - PRO2091/Tray - PRO2092)
Steam Pre-vacuum
270 - 272°F (132 - 133°C)
4 minutes
8 minutes minimum *
Steam Gravity
270 - 272°F (132 - 133°C)
15 minutes
8 minutes minimum *
Steam Gravity
250 - 254°F (121 - 123°C)
40 minutes
8 minutes minimum *
Sterilization Guidelines for Sterilizing PowerPro Attachments in the PowerPro Complete
Sterilization Case (PRO2090)
Steam Pre-vacuum
270 - 272°F (132 - 133°C)
4 minutes
10 minutes minimum *
Steam Gravity
270 - 272°F (132 - 133°C)
15 minutes
10 minutes minimum *
Steam Gravity
250 - 254°F (121 - 123°C)
40 minutes
10 minutes minimum *
* CAUTION: The recommended dry cycle must be run on all attachments every time they
are sterilized. Failure to use a dry cycle may lead to reduced product performance or prema-
ture product failure. Operation of an attachment that is not completely cool and dry may
decrease performance and/or reliability.