3.1.2
Attachment Cleaning
Instructions
Clean attachments as soon as possible after use.
1.
Remove the attachment from the handpiece
prior to cleaning.
2.
Thoroughly scrub attachments with a clean,
soft brush dampened with a mild,
pH-balanced detergent. Remove all traces
of blood, debris and stains.
Do Not
immerse
equipment in soap solution or
rinse water.
3.
Clean the cannulated section of the attach-
ment by performing the following:
(a) Feed the wire end of a cleaning brush
through the attachment.
(b) Pull the brush completely through the
attachment. Repeat until all debris is
removed.
4.
Manipulate all moving parts of attachments
to ensure all debris is removed. If not, clean
again until all debris is removed.
5.
Rinse attachments under running water to
remove all traces of soap.
6.
Flush all surfaces free of tap water with dis-
tilled water to prevent metal discoloration.
7.
Gently shake attachments free of water and
wipe the surfaces with a clean, lint-free
towel.