For Non-Manifold Use:

 

If  purchased  with  diaphragm  valves,  the  male  ½  NPT  ends  of  the  diaphragm  valve  elbows  are 
installed into the inlet ports of the switchover system. Install the male ¼ NPT end of each pigtail into 
the ¼ NPT ports of the diaphragm valves. The connection available at the other end of the pigtail 
will  depend  on  how  the  system  was  purchased.  If  the  pigtail  does  not  have  a  cylinder  connection 
installed, there will be a female ¼ NPT thread on the end of each pigtail. The female ¼ NPT pigtail 
end  will  connect  to  the  user’s  system.  If  configured  with  a  CGA  connection  or  foreign  inlet 
connection,  the  connection  will  be  specific  to  the  particular  cylinder  of  gas  to  be  used.  Some 
connections do require the use of a gasket. Please note that the material of the gasket must also be 
compatible with the type of gas being used. Be familiar with the type of connection being used, and 
its proper procedures for installation.

 

 
 
 
 

 
 

Figure 8 - Pigtail

 

 

 

 
 
 
 

 

 

Connecting to a Cylinder:

 

Figure 9 -  Diaphragm Valve with

 

1/2” NPT Street Elbow

 

1.  Before removing the cylinder cap, move the cylinder of gas to the work site:

 

a.  Secure cylinder to the floor, wall, or bench with appropriate chain, strap, or stand to 

prevent toppling.

 

b.  Remove the cylinder cap.

 

c.  Be sure the cylinder valve is tightly closed (clockwise).

 

d.  Remove the cylinder valve plug, if any.

 

e.  Inspect the cylinder valve and threads for damage or contamination.

 

 

2.  Secure the cylinder connection to the cylinder in the following manner:

 

a.  Do not force. Tightening the nut onto the cylinder connection should be easy. If it is not, 

the connection may be wrong for the type of gas being used.

 

b.  Left-hand threads are used on some cylinder connections. A notch in the middle of the hex 

nut typically indicates a left-hand thread.

 

c.  Gaskets are used on some inlet connections. Be sure the gasket is in good shape. Do not 

over-tighten to avoid squashing the gasket into the gas line. Keep extra gaskets on hand.

 

d.  Never use oil or grease on regulator or cylinder fittings, as it may contaminate pure 

gases or create a fire hazard.

 

 

Installing the Outlet Connection:

 

The  AutoS

witch  system  has  the  outlet  connection  at  the  top  of  the  box.  The  connection  is  a  1/4” 

tube  connection.  Connect  tubing  to  the  AutoSwitch  that  is  compatible  with  the  application  and  is 
capable of handling the pressure required.

 

 

-8-

 

Summary of Contents for 536 Series

Page 1: ...s AutoSwitch Systems For Use with Liquid Cylinders and High Pressure Gases in High Purity Applications INSTALLATION AND OPERATION INSTRUCTIONS Controls Corporation of America 1501 Harpers Road Virgini...

Page 2: ...THIS PAGE INTENTIONALLY LEFT BLANK 2...

Page 3: ...imary bank falls below this value gas will be pulled from the reserve side Models of these AutoSwitches are available with or without pressure switch gauges Models with pressure switch gauges can be c...

Page 4: ...ifold Always leak test any manifold or distribution pipeline before using Always be sure that the gas in a pipeline is the correct gas for the intended use Always install a blind gland and nut or pipi...

Page 5: ...use of carbon dioxide 13 C G A Pamphlet G 6 1 Standard for Low Pressure Carbon Dioxide Systems at Consumer Sites INSTALLATION The site chosen for the installation of the system shall be level well ve...

Page 6: ...oint extending approximately 10 inches to the left and right of center 3 Remove the mounting bracket from the rear of the AutoSwitch console by removing the retaining nuts at the bottom of the mountin...

Page 7: ...ad sealant that remains on the threads d Start the PTFE tape on the first thread leaving a slight section of the chamfer exposed as shown in Fig 6 Make sure the tape does not overlap the end of the fi...

Page 8: ...the cylinder of gas to the work site a Secure cylinder to the floor wall or bench with appropriate chain strap or stand to prevent toppling b Remove the cylinder cap c Be sure the cylinder valve is ti...

Page 9: ...ns if compatible with the application which indicates leaks by bubbling To further check for leaks or if the leak detection solution cannot be used close the cylinder valve for a period of time recomm...

Page 10: ...e switchover point of the AutoSwitch The alarm is used to provide an audible and visual warning that a changeover is about to occur If the nearly depleted cylinders are changed when the alarm LEDs ill...

Page 11: ...cylinders simultaneously pressure is decreasingonboth inletgaugesatthesametime 1 Observe the inlet pressure It may be necessary to do this during times of the heaviest use If the inlet pressure is le...

Page 12: ...rvice Please contact a Customer Service Representative at 1 800 225 0473 to receive this number You will be asked to provide 1 Model Number 2 Gas Service 3 Inlet pressure and type of gas supply 4 Outl...

Page 13: ...THIS PAGE INTENTIONALLY LEFT BLANK 13...

Page 14: ...THIS PAGE INTENTIONALLY LEFT BLANK 14...

Page 15: ...ther than CONTROLS CORPORATION OF AMERICA or if the equipment has been subject to abuse misuse negligence or accident CONTROLS CORPORATION OF AMERICA s sole and exclusive obligation and Buyer s sole a...

Page 16: ...ADI 3198 D Certified ISO 9001 Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 Telephone 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com...

Reviews: