INLET AND OUTLET CONNECTIONS

   

General:

 

Use  an  open-end  wrench,  not  a  pipe  wrench,  to  install  accessories  to  the  AutoSwitch  system. 

ALWAYS  hold  the  bulkhead  fitting  with  a  wrench  when  installing  the  accessories  to  the 
AutoSwitch system. Do not allow the bulkhead fitting to turn; otherwise, connections inside 
the box may be loosened.

 NPT connections require the use of PTFE tape on the threads to make 

a  gas  tight  seal  (see  Figure  6).  CONCOA  uses  PTFE  tape  on  all  of  its  NPT  connections.  Follow 
these rules when using PTFE tape.

 

 

a. Do not use oil or grease on fittings. This is especially important for oxidizing gas service. 

 

b. PTFE tape must be used on NPT threads to ensure a gas tight seal. Avoid impinging on the gas stream.  

 

 

On stainless steel connections, PTFE tape also helps to prevent the connections from galling together  

 

 

when tightening or loosening. 

 

c. Before applying PTFE tape, inspect the NPT threads, and, if necessary, clean the fitting to remove any dirt  

 

 

or thread sealant that remains on the threads.  

 

d. Start the PTFE tape on the first thread leaving a slight section of the chamfer exposed as shown in Fig. 6.  

 

 

Make sure the tape does not overlap the end of the fitting.  

 

e. As tape is wound in the direction of the thread spiral, pull tightly on the end of the tape so that the tape  

 

 

conforms to the thread.  

 

f. Apply at least 2 but no more than 3 layers of tape to the threads, and cut off excess tape.  

 

g. Press the end firmly into the threads.  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Figure 6 - Pictorial View of Tape Seal 

 

Installing the inlet connection:

 

Warning: The left bank of the AutoSwitch system has a maximum working pressure of  
   

600 PSI (41 bar). Do not install high pressure cylinders to the left bank of the  

   

AutoSwitch system.

 

When using a Maniflex system:

 

The  ½  male  NPT  end  of  the  manifold  connector  will  need  to  be  installed  in  the  inlet  port  of  the 
AutoSwitch.  The  other  end  of  the  manifold  connector  will  be  connected  to  the  starter  block  of  the 
528 series Maniflex system. Please follow the instructions provided with the Maniflex system when 
connecting and operating the manifold system.

 

 
 
 
 
 

Figure 7 - Manifold connector for the Maniflex system 

 

-7-

 

1/2” NPT M 

1/2” NPT M 

Summary of Contents for 536 Series

Page 1: ...s AutoSwitch Systems For Use with Liquid Cylinders and High Pressure Gases in High Purity Applications INSTALLATION AND OPERATION INSTRUCTIONS Controls Corporation of America 1501 Harpers Road Virgini...

Page 2: ...THIS PAGE INTENTIONALLY LEFT BLANK 2...

Page 3: ...imary bank falls below this value gas will be pulled from the reserve side Models of these AutoSwitches are available with or without pressure switch gauges Models with pressure switch gauges can be c...

Page 4: ...ifold Always leak test any manifold or distribution pipeline before using Always be sure that the gas in a pipeline is the correct gas for the intended use Always install a blind gland and nut or pipi...

Page 5: ...use of carbon dioxide 13 C G A Pamphlet G 6 1 Standard for Low Pressure Carbon Dioxide Systems at Consumer Sites INSTALLATION The site chosen for the installation of the system shall be level well ve...

Page 6: ...oint extending approximately 10 inches to the left and right of center 3 Remove the mounting bracket from the rear of the AutoSwitch console by removing the retaining nuts at the bottom of the mountin...

Page 7: ...ad sealant that remains on the threads d Start the PTFE tape on the first thread leaving a slight section of the chamfer exposed as shown in Fig 6 Make sure the tape does not overlap the end of the fi...

Page 8: ...the cylinder of gas to the work site a Secure cylinder to the floor wall or bench with appropriate chain strap or stand to prevent toppling b Remove the cylinder cap c Be sure the cylinder valve is ti...

Page 9: ...ns if compatible with the application which indicates leaks by bubbling To further check for leaks or if the leak detection solution cannot be used close the cylinder valve for a period of time recomm...

Page 10: ...e switchover point of the AutoSwitch The alarm is used to provide an audible and visual warning that a changeover is about to occur If the nearly depleted cylinders are changed when the alarm LEDs ill...

Page 11: ...cylinders simultaneously pressure is decreasingonboth inletgaugesatthesametime 1 Observe the inlet pressure It may be necessary to do this during times of the heaviest use If the inlet pressure is le...

Page 12: ...rvice Please contact a Customer Service Representative at 1 800 225 0473 to receive this number You will be asked to provide 1 Model Number 2 Gas Service 3 Inlet pressure and type of gas supply 4 Outl...

Page 13: ...THIS PAGE INTENTIONALLY LEFT BLANK 13...

Page 14: ...THIS PAGE INTENTIONALLY LEFT BLANK 14...

Page 15: ...ther than CONTROLS CORPORATION OF AMERICA or if the equipment has been subject to abuse misuse negligence or accident CONTROLS CORPORATION OF AMERICA s sole and exclusive obligation and Buyer s sole a...

Page 16: ...ADI 3198 D Certified ISO 9001 Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 Telephone 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com...

Reviews: