Section 3 

Operation 

 
This section describes how to set up and make measurements with your tester.  We recommend 
that you read the entire section carefully so that you can use all of the features of your tester. 
 

Setting up your Tester 

 
Your  tester  is  shipped  in  a  special  protective  container  that  should  prevent  damage  to  the 
instrument during shipping.  Check the shipping order against the contents of the container and 
report  any  damage  or  short  shipment  to  Compliance  West  USA.    The  container should include 
the following: 
 

 

The MegaPulse 1.2x50-16PF PV Tester 

 

Two large alligator clips with insulating boots (one red, one black) 

 

An 18 AWG Line Power Cord 

 

This Instruction Manual 

 

Calibration certificate with waveform calibration pictures 

 
If  reshipment  of  the  instrument  is  necessary,  please  use  the  original  shipping  container.    If  the 
original  shipping  container  is  not  available,  be  sure  that  adequate  protection  is  provided  to 
prevent damage during shipment.  We recommend that the instrument be surrounded by at least 
three inches of shock-absorbing material on all sides of the container. 
 
Remove the tester from its container and place it on a test bench.

  CAUTION: The MegaPulse tester 

is very heavy.  Two-person lift is required. 

 

AC Line Voltage Requirements 

 
AC line voltage requirements for your Tester are noted on the rear panel of the instrument.  Do 
not  connect  the  instrument  to  a  different  voltage  source.    The  cord  packaged  with  your 
MegaPulse Tester is for use in the United States.  If another power cord must be used, the cord 
must  be  rated  for  the  maximum  current  noted  on  the  rear  panel.    It  must  also  meet  the 
requirements  of  IEC  227  or  IEC  245,  and  mains  cords  that  are  certified  or  approved  by  any 
recognized national test house are regarded as meeting this requirement.   

 

Summary of Contents for MegaPulse

Page 1: ...MegaPulse IMPULSE TESTER 1 2x50 16PF PV Instruction Manual...

Page 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 3: ...3...

Page 4: ...Testing Area 6 Safety Techniques 7 Using the MegaPulse Impulse Tester 7 Introduction 9 Operation 11 Setting up your Tester 11 AC Line Voltage Requirements 11 Fuse Replacement 12 Front and Rear Panel...

Page 5: ...5...

Page 6: ...re that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especial...

Page 7: ...er test from the MegaPulse until the front panel meter reads a safe level 40 volts or less is generally considered a safe level Pressing the TRIGGER button before disconnecting main power or turning t...

Page 8: ...5 Ohms and the MegaPulse is capable of delivering thousands of Amps of current into a short circuit While the MegaPulse can withstand a short circuit at the output terminals it should NOT be used in t...

Page 9: ...Discharge progress is shown on the front panel meter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Output Peak 800V 16kV To...

Page 10: ...follow test sequences varying test voltage and time between tests and keep a record with test information Compatible with Windows 32bit platform 220 230 or 240 If a different line voltage is desired...

Page 11: ...e use the original shipping container If the original shipping container is not available be sure that adequate protection is provided to prevent damage during shipment We recommend that the instrumen...

Page 12: ...h to the O or off position 2 Unplug the instrument from the source of supply 3 Remove the power inlet cord from the instrument 4 Using a small screwdriver pry open the fuse holder door 5 Replace the f...

Page 13: ...13 Figure 3 Controls Indicators Connectors MegaPulse 1 2x50 16PF PV Front Panel...

Page 14: ...been pressed 5 CHARGE indicator This Yellow indicator is lit to show that pressing the CHARGE switch is the next logical step in a test sequence Pressing the Charge switch causes the CHARGE indicator...

Page 15: ...ied cordset to connect the MegaPulse tester to appropriate source of supply Fuse holder provides access for Fuse replacement and the Power Switch is used to turn the tester ON and OFF 2 Fuse replaceme...

Page 16: ...setup it is possible to see very large common mode noise which can be mis interpreted as high frequency noise on the front edge of the impulse waveform Figure 5 shows a test setup to measure common m...

Page 17: ...17 Figure 5 Common mode noise measurement setup Figure 6 Common mode noise Figure 7 Common mode noise superimposed on the output waveform...

Page 18: ...nds on the front meter Turn Clockwise to increase the setting Voltage Setting Point 7 Push the yellow CHARGE button Verify the red TRIGGER indicator is now lit 8 After the front panel meter indicates...

Page 19: ...voltage is present on the main storage capacitor even when the tester is first turned ON This is due to inherent charging of the internal capacitors Pressing the TRIGGER switch will discharge the cap...

Page 20: ...n Information The Megapulse 1 2x50 16PF PV has been fully calibrated at the factory in accordance to our published specifications It is recommended that you have this instrument re calibrated and safe...

Reviews: