2
Section 1
Introduction
This manual contains complete operating, maintenance and calibration instructions for the Compliance
West USA Model GFM-200A Ground Bound Tester.
The instrument is a bench-type ground circuit tester.
The GFM-200A features the variac knob, current and voltage meter, always operate with precaution.
The GFM-200A meets all safety agency criteria for ground bound testers.
Your tester is warranted for a period of one year upon shipment of the instrument to the original
purchaser.
Specifications
Specifications for the GFM-200A are listed in Table 1-1.
GFM-200A Specifications
ELECTRICAL
Output
0-200 Amps ac
±
3% into any load - short circuit to
12.5 milliohms
Voltage
5 volts max. open circuit
Meters
0-6 V ac,
±
3% full scale
0-200A ac,
±
3% full scale
ENVIRONMENTAL
Operating Temperature
15-40
°
C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
90-127 or 180-254 volts, 50/60 Hz
Dimensions
11.250(W) x 4.250(H) x 12.000(L) in.
Weight
30 lbs
Table 1-1. GFM-200A Specifications