8-2
8.
Troubleshooting
8.3 Reset
The CMM has a reset button that simulates a power on/off
cycle.
The primary use is after a program has been
loaded. However, it may be used at any time. It does not
reset the CMM board to factory settings.
fig 8.3-1 Reset Button
8.4 Factory Menu
The factory menu may be accessed to re-initialize or
change the date/time screen.
To access the Factory Menu press Menu 9 684075 at the
main menu. During key entry the display will not change
until the last digit is entered.
1) CONFIG
2) INIT
3) TEST
4)TIME
5) SERIAL #
fig 8.4-1
8.4.1 CONFIG Menu
The menu will allow the user to reconfigure the CMM unit
for the two options: CO and MO or CO only.
Press 1 to enter configuration screen
CONFIG =
19
NEW CONFIG =
FIG 8.4.1-1
The first line indicates the current configuration. The user
may enter the new configuration and press enter.
The two configurations are:
19 = Carbon Monoxide and Moisture
5 = Carbon Monoxide only
8.4.2 Initialization Menu
This menu will initialize the CMM and reset it to
factory
settings
and
clear
the
E-PROM
memory.
This should only be done by
authorized personnel.
CMM units with an asterisk in the version
number on the main screen indicates that an
initialization is required.
Initialization is also needed after a new version
program is entered into the CMM unit.
Press 2 in the factory menu to enter the
initialization menu.
PRESS
9
TO
RESET HOURS & EEPROM
fig 8.4.2-1
Press 9 to confirm the initialization must be
performed.
This menu will test the
LEDs for operation.
Once started the LEDs will flash on-off for
approximately one minute. The CMM may be
reset by turning the unit off-on or by pressing
the reset button.
8.4.3 Test Menu
PRESS 9 TO TEST
fig 8.4.3-1
Press 9 to confirm entry into the test sequence.