50
IE-LGR-UxxxxM-22
Connection:
Pluggable terminal block, maximum wire cross section
1.5 mm
2
. Recommended length of the Pt1000 probe
cables is maximum 15 m, a length of 30 m should not
be exceeded. Maximum length of the binary inputs
cables is 30 m. It is recommended to use shielded
cables.
CAUTION: The inputs are not galvanically isolated!
Response time:
Is determined by the response time of the used probe (it
valid if Measuring interval is set to 1 s).
Resolution:
0.1 °C converter range: 16 bits
Recommended calibration interval:
2 years
U3120M
________________________________________
Measured values:
Internal temperature, relative humidity.
Dew-point temperature calculated from the internal
temperature and relative humidity.
Ranges:
Temperature: (-
20 to +60) °C
Relative humidity: (0 to 100) %RH without permanent
condensation.
Dew-point temperature: -60
°C to +60 °C
Accuracy
(*1)
:
Temperature: ± 0.4 °C
Relative humidity:
-
sensor accuracy ±1.8 %RH
at 23 ºC
in the
range of 0 to 90 %RH
-
hysteresis < ±1 %RH
- non-
linearity < ±1 %RH
- temperature error: see diagrams in
Dew-point temperature:
±1.5 °C at an ambient
temperature of T< 25
°C and RH >30 %.
For details see diagrams in
Response time (air flow 1 m/s approx.)
(*2)
:
Temperature: t
63
< 2 min, t
90
< 8 min (temperature step
20 °C)
Relative humidity: t
63
< 45 s, t
90
< 4 min, t
99
<4 h.
(Measured at and constant temperature for
30% humidity step)
Summary of Contents for U0110M
Page 5: ...IE LGR UxxxxM 22 5 Datalogger UxxxxM U0141M model...
Page 60: ...60 IE LGR UxxxxM 22 Dimensions U0110M U8410M U0121M...
Page 61: ...IE LGR UxxxxM 22 61 U0141M U3120M U4440M...
Page 62: ...62 IE LGR UxxxxM 22 U3121M U3631M...
Page 63: ...IE LGR UxxxxM 22 63 U0141TM U0843M U5841M U6841M U7841M...