6
IE-LOG-S3121P-03
Reading on the display in usual operation (logger switched ON)
After switching ON the logger all LCD symbols are displayed for checking
the display.
Then actual date and time in logger is displayed for approximately 4 s.
Consequently reading of estimated remaining battery capacity for
approximately 2 s is displayed (values 0 to 100%).
If display is switched ON
, actual reading of measured values is displayed -
ambient temperature (°C) on the LCD upper line, relative humidity (%RH)
on the LCD lower line. Symbol LOG indicates data logging in progress – if
it blinks, data memory is filled up to more than 90%.
Every 5 s display is automatically switched to displaying of other measured
or calculated quantity.
Logger now displays ambient temperature and dew point temperature (LCD
line marked by symbol DP).
Switched ON logger permanently (with 10 s interval) updates memory of
minimum and maximum values of each measured (or calculated) quantity. If
displaying of MIN/MAX values is selected, minimum measured values are
displayed step by step (indicated by symbol MIN) and then similarly
maximum measured values of all quantities (indicated by symbol MAX).
The entire cycle is repeated periodically, i.e. reading of actual measured
values follows.
If display is switched OFF,
all above readings are displayed up to
estimated remaining battery capacity and then display goes out. If logger is
switched ON symbol LOG is displayed (it blinks if memory occupation is
higher than 90%).
If display is OFF and logger is in mode when record runs only when alarm is
active, the LOG symbol is replaced by adjacent symbol "–" (hyphen). It
appears in case, all measured values are inside of adjusted alarm limits and
data logging therefore does not run. Displayed symbol indicates logger is
ON.
If information on actual measured values is needed, it is possible anytime to display reading
display by means of the magnet (only if communication adapter is not connected permanently). Plug
magnet into guide slots from logger front side for approximately 4 s and wait till reading on the
display appears. If logger has enabled the function switch OFF by magnet, do not remove magnet
from guide slots before decimal point symbol goes out – logger would be switched OFF! Display
reading initialized by magnet goes out automatically after 30 s. Remove magnet from slots anytime
during actual reading is ON or later.