  IE-LOG-S3121P-03 

Communication with computer: via RS232 (serial port) by means of COM adapter or USB port 

by means of USB adapter; data transfer from logger via communication adapter is optical 

Real time clock: adjustable from computer, integrated calendar including leap years 
Error of internal RTC: < 200 ppm (i.e. 0.02 %, 17.28 s in 24 h) 
Power: Lithium battery 3.6 V size AA 
Typical battery life: 2,5 years 
Protection: electronics IP67, sensors are protected by a cover with IP40 protection 
Operational conditions: 

Operational temperature range: instrument -30 to +70 °C, external probe –30 to +105°C 
Operational humidity range: 0 to 100 %RH 
Operational pressure range of the probe: up to 25 bar 
Air flow velocity: up to 25 m/s at a pressure of 1 bar  (1m/s at a pressure of 25 bar) 
Specification of outer characteristics accordingly to Czech National Standard 33-2000-3: 

normal environment accordingly to appendix NM:  AE1, AN1, AR1, BE1 

Operational position: negligible  
Logger installation: by self adhesive Dual Lock, applied to clean, flat surface 
Not  allowed  manipulation:  it  is  not  allowed  to  remove  sensor  cover  and  mechanically 

damage sensor under cover. Sensors of temperature and humidity should not get to 
direct contact with water or other fluids. 

Limit condition – logger without probe: temperature -40 to +70 °C, humidity 0 to 100 %RH 
Storing condition: temperature -40 to +85 °C, humidity 0 to 100 %RH 
Mechanical connection of the probe: G1/2 with O-ring 
Dimensions of the logger without connector and probe: 93 x 64 x 29 mm 
Dimensions of the external probe: see dimensional chart (default cable length 1 meter) 
Weight including battery without probe: approximately 115 g 
Material of the case: ABS 
Material of the probe: duralumin with the black eloxal surface finish 

 
 
 
 

Logger operation 

Logger  comes  complete  with  installed 

battery and switched OFF. Before operation 
it is necessary by means of installed user PC 
software to set logging parameters and other 
features.  Free  user  program  is  available  to 
download  at  www.cometsystem.cz  .  For 
communication 

with 

the 

PC 

communication  adapter  is  necessary  (not 
included  in  delivery).  For  connection  via 
RS232 serial port it is necessary to use COM 
ADAPTER,  for  connection  via  USB  port  it 
is  necessary  to  use  USB  ADAPTER. 
Connect  adapter  connector  to  proper 
computer  port  and  plug  the  adapter  to  the 
guide slots on the side of logger.  

Connecting  logger  to  the computer  by  means  of 
communication adapter 

 

Notice:

 USB connector can be located also from the computer front side 

 

Summary of Contents for LOGGER S3121P

Page 1: ...LOGGER S3121P Temperature and relative humidity logger with external probe for compressed air measurement Instruction Manual...

Page 2: ...led for each quantity individually Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging...

Page 3: ...0 to 105 C Resolution 0 1 C Accuracy 1 5 C at ambient temperature T 25 C and relative humidity RH 30 for more details see graphs below Response time with stainless steel mesh cover air flow approximat...

Page 4: ...s of temperature and humidity should not get to direct contact with water or other fluids Limit condition logger without probe temperature 40 to 70 C humidity 0 to 100 RH Storing condition temperature...

Page 5: ...ary Logging interval between to subsequent measurements is specified by the user Memorizing of the first value is synchronized with the internal real time clock so the logging is performed at sharp mu...

Page 6: ...maximum measured values of all quantities indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings...

Page 7: ...essages displayed on the LCD beyond usual operation If measured value is out of the measurable or displayable range numeric reading is replaced by hyphens If memory is totally full filled in non cycli...

Page 8: ...e magnet from guide slots and logger switches ON Switching the logger OFF by magnet The procedure is identical with the above procedure for switching ON If decimal point does not appear after 1 s it i...

Page 9: ...ger Attention inserted battery with incorrect polarity causes logger damage put the lid back again and screw four screws Be careful to have the rubber sealing properly in the slot and tighten the scre...

Page 10: ...rrupting the main line closing valve 5 the sample flow is regulated by this valve Measurement accuracy is typically not affected by the sample flow rate but at low speeds increasing response time outl...

Reviews: