background image

IE-LOG-S3121P-03 

 

  3 

Technical parameters: 

 

 

Measuring parameters: 

  Temperature (RTD sensor Pt1000/3850ppm): 

 

Measuring range: -30 to +105 °C 

   

 

Resolution: 0.1 °C 

   

 

Accuracy: ± 0.4 °C from –30 to +80°C, ± 0.5 °C over +80°C 

  Relative humidity (reading is temperature compensated at entire temperature range): 

 

Measuring range: 0 to 100 %RH 

   

 

Resolution: 0.1 %RH 

   

 

Accuracy: ± 2.5 %RH from 5 to 95 %RH at 23 °C 

  Dew point (value calculated from temperature and humidity): 

 

   

 

Range: -60 to +105 °C 

   

 

Resolution: 0.1 °C 

 

 

 

Accuracy: ±1.5 °C at ambient temperature T<25ºC and relative humidity  

 

 

 

 

RH>30 %, 

for more details see graphs below. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Response time with stainless steel mesh cover (air flow approximately 1 m/s): 

  air temperature: t90 < 16 min (temperature step 20 °C) 
  relative humidity: t90 <  30 s (humidity step 65 %RH, constant temperature) 

Measuring interval, alarm evaluation, MIN/MAX memory update and display refresh:  

default setting: 10 s 
low power mode: 1 min 

Logging interval: 

default setting: 10 s to 24 h 
low power mode: 1 min to 24 h 

Total memory capacity:  

for non cyclic mode 

16 252 T+RH readings 

for cyclic mode 

15 296 T+RH readings 

Summary of Contents for LOGGER S3121P

Page 1: ...LOGGER S3121P Temperature and relative humidity logger with external probe for compressed air measurement Instruction Manual...

Page 2: ...led for each quantity individually Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging...

Page 3: ...0 to 105 C Resolution 0 1 C Accuracy 1 5 C at ambient temperature T 25 C and relative humidity RH 30 for more details see graphs below Response time with stainless steel mesh cover air flow approximat...

Page 4: ...s of temperature and humidity should not get to direct contact with water or other fluids Limit condition logger without probe temperature 40 to 70 C humidity 0 to 100 RH Storing condition temperature...

Page 5: ...ary Logging interval between to subsequent measurements is specified by the user Memorizing of the first value is synchronized with the internal real time clock so the logging is performed at sharp mu...

Page 6: ...maximum measured values of all quantities indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings...

Page 7: ...essages displayed on the LCD beyond usual operation If measured value is out of the measurable or displayable range numeric reading is replaced by hyphens If memory is totally full filled in non cycli...

Page 8: ...e magnet from guide slots and logger switches ON Switching the logger OFF by magnet The procedure is identical with the above procedure for switching ON If decimal point does not appear after 1 s it i...

Page 9: ...ger Attention inserted battery with incorrect polarity causes logger damage put the lid back again and screw four screws Be careful to have the rubber sealing properly in the slot and tighten the scre...

Page 10: ...rrupting the main line closing valve 5 the sample flow is regulated by this valve Measurement accuracy is typically not affected by the sample flow rate but at low speeds increasing response time outl...

Reviews: