Section 4 Operation
Fluids Type
is used to identify the type of fluid used. The controller uses the fluid type to automati
-
cally set certain operating parameters.
Pump Adjustment
is used to review/set the
desired pump speed.
Water
EG-Water
PG-Water
Other
Menu
Low
High
Menu
2.
With the desired fluid selected press
, or use
the
Menu
window, to save and return to
the previous display.
1.
With
Fluid Type
highlighted, press
to
display the list of acceptable fluids.
Highlight the desired fluid and press
to
select it.
1.
With
Pump Adjustment
highlighted, press
to display the speeds.
Highlight the desired speed and press
to
select it.
Fluid system limits High °C
Low °C
All units:
Water
+98
+2
EG-Water
+103
-33
PG-Water
+103
-33
Controller 1 units:
Other
+103
-33
Controller 2 units:
Other
+103
-33
SIL 100
+75
-30
SIL 180
+150
-30
SIL 300
+150
+80
SYNTH 60
+45
-30
SYNTH 200
+150
+30
SYNTH 260
+150
+45
Controller 3 units:
Other
+200
-90
SIL 100
+75
-75
SIL 180
+200
-40
SIL 300
+200
+80
SYNTH 60
+45
-50
SYNTH 200
+200
+30
SYNTH 260
+200
+45
NOTE
The range is also limited by the bath
temperature range, see Section 2.