Section 3 Installation
3-3
Cole Palmer
CAUTION
Ensure all communication and electrical connections are made
prior to starting the unit.
Ensure the line cord from the bath is securely connected to the rear
•
of the controller.
Ensure the communications cable from the bath is connected to the
•
RJ45 connectors (similar to Ethernet) on the rear of the controller.
Connect the supplied line cord to the controller and to a properly
•
grounded outlet.
Cable connection required
for proper communication
between the bath and the
circulator.
1/2 Immersion Circulator
Communications cable
from bath
Line cord from bath
Line cord connection
I/O Circuit Protector
Return Flow
Outlet Flow
3 Immersion Circulator
6200 Remote
Temperature Sensor
6200 USB
Connection
Return Flow
Outlet Flow
1 2 3 4 5
6 7 8 9
Never apply line voltage to the connection.
The optional remote temperature sensor is enabled using the controller,
see Section 4.
Pin
1
White
2, 3 NA
4
White
5, 6 NA
Pin
7
Red
8
NA
9
Red (4th wire not connected to the control board)
WARNING
Remote Temperature
Sensor (3 Only)