REMOTE CW/CCW: This input allows the user to change the direction
of flow remotely.
REMOTE START/STOP: When the START/STOP
input
is open, the
drive can still be started using the START/STOP key, PRIME key, or
PRIME input. In remote modes the drive will also run if there is sufficient
current or voltage at the input.
Closing the START/STOP
input
will cause the drive to run until the
START/STOP
input
opens or the START/STOP key is pressed. In Copy
dispense mode, only a momentary START/STOP closure is needed to start
the drive. If the drive is already running in one of the dispense modes, a
momentary START/STOP closure will stop the drive. In SET COPY
MODE, the START/STOP
input
functions the same as in
CONTINUOUS MODE; closing it will cause the drive to run until it
opens.
REMOTE PRIME: This input allows the user to prime the pump
remotely.
REMOTE VOLTAGE or CURRENT INPUT ENABLE: This input
enables the user to remotely activate the Voltage or Current input.
Enabling this will turn ON the last known input (Voltage or Current) that
had been activated from the Remote control menu and saved.
Remote Control Inputs
and Outputs
(continued)
MASTERFLEX
®
Digital Pump Drive Operating Manual 3-23