This will take you to the setpoint screen. The user has an option to adjust
the minimum and maximum setpoints for voltage and RPM. The setpoints
being edited will be blinking on the slope. To change which setpoint to
edit use the UP or DOWN to select screen. By default the minimum
(MIN) voltage is set to 0.1V DC and the RPM is set to 0.0. The
maximum (MAX) is set to 10.0V DC and the RPM is set to maximum.
The scaling can be inverted, if necessary. After both setpoints have been
adjusted, save both setpoints from the upper setpoint setting screen by
highlighting the save icon and pressing ENTER.
To confirm VOLTAGE INPUT MODE is ON, press the BACK key until
the CONTINUOUS PUMP MODE screen appears. The Voltage input
icon will be displayed.
NOTE:
When Voltage Input is selected the drive will not start until the
VOLTAGE INPUT is ON and in CONTINUOUS PUMP MODE.
Remote Control
Voltage Input Menu
(continued)
3-14
MASTERFLEX
®
Digital Pump Drive Operating Manual
Cole-Parmer
Section 3
Operation