![Cole Parmer EH-200 Series Instruction Manual Download Page 11](http://html1.mh-extra.com/html/cole-parmer/eh-200-series/eh-200-series_instruction-manual_2635365011.webp)
11
8.
TECHNICAL SPECIFICATION
8.1.
General
specification
Mains input supply voltage (115V~ AC)
115V – AC ± 10% at 50/60Hz
Mains input supply voltage (230V~AC)
230V – AC ± 10% at 50/60Hz
Plug and Lead set
3 core 2 meters long.
(No cable supplied on EH-200-6-L-115)
Operating Ambient Temperature
Up to 40 °C.
Heating Element Construction
Thermal insulated element wire
stitched into a cartridge construction
Heating Element Temperature
800°C Max
Case construction
Stainless-steel
Thermal Insulation
Ceramic Fibre / Mineral wool
Clamps for Support rods
12.7mm dia max
8.2.
Power consumption and fuse rating
Unit
Capacity
Total power
Fuse rating
EH-200-2
100-300
600
F5A 6.35 x 32mm Glass Quick Blow
EH-200-2-115
100-300
600
F7A 6.35 x 32mm Ceramic
EH-200-2-EU
100-300
600
F5A 6.35 x 32mm Glass Quick Blow
EH-200-2-L
500-800
1100
F7A 6.35 x 32mm Ceramic
EH-200-2-L-115
500-800
1100
F15A 6.35 x 32mm Glass Quick Blow
EH-200-2-L-EU
500-800
1100
F7A 6.35 x 32mm Ceramic
EH-200-6
100-300
1800
F5A 6.35 x 32mm Glass Quick Blow
EH-200-6-115
100-300
1800
F10A 6.35 x 32mm Glass Quick Blow
EH-200-6-EU
100-300
1800
F5A 6.35 x 32mm Glass Quick Blow
EH-200-6-L
500-800
3800
F10A 6.35 x 32mm Glass Quick Blow
EH-200-6-L-115
500-800
3800
See note**
EH-200-6-L-EU
500-800
3800
F10A 6.35 x 32mm Glass Quick Blow
**Note:
35Amp fuses to be
fi
tted at the supply/distribution point see section 5
8.3.
Weight and Dimensions
EH-200-2, EH-200-2-L
Weight 2.0kg
EH-200-6, EH-200-6-L
Weight 5.8kg
The Ingress Protection rating for this product is classified as IPX 0.