background image

Introduction ____________________________________________________________________ 2
General safety instructions ________________________________________________________ 3

Dangers associated with the operation  ___________________________________________________________ 3
Photobiological safety LED, important safety instructions  ____________________________________________ 3
Prevention of biological and infectious hazards  ____________________________________________________ 3
Disinfection and decontamination:   ______________________________________________________________ 4
Model with rechargeable batteries  _______________________________________________________________ 5
Environment, storage and use ___________________________________________________________________ 5

Models  _________________________________________________________________________ 6
Components of the microscope  ____________________________________________________ 7
Preparing the microscope for use  __________________________________________________ 8

Assembling steps  _____________________________________________________________________________ 8

Operation  ______________________________________________________________________ 9

Setting up the illumination  _____________________________________________________________________ 9
Placing the specimen slide ______________________________________________________________________10
Focusing and slide protection  ___________________________________________________________________10
Adjusting the focusing tension  __________________________________________________________________10
Eyepieces  ____________________________________________________________________________________10
Abbe condenser  ______________________________________________________________________________11
The field (Köhler) diaphragm ____________________________________________________________________11
Adjusting the aperture diaphragm  _______________________________________________________________11
Use of the S100x oil-immersion objective __________________________________________________________12
Safety device _________________________________________________________________________________12
Illumination __________________________________________________________________________________12

Maintenance and cleaning  _______________________________________________________ 12

Cleaning the optics  ____________________________________________________________________________12
Maintenance of the stand  ______________________________________________________________________13
Changing the batteries _________________________________________________________________________13
Replacing the fuse  ____________________________________________________________________________13

Accessories  ____________________________________________________________________ 13
Spare Parts  ____________________________________________________________________ 13
Warranty ______________________________________________________________________ 13

Contents

Introduction 

Thank you for purchasing the Cole-Parmer MSU-200 Series compound binocular microscope. These entry-level upright 
microscopes are developed for use at schools and laboratories. They offer corded and cordless battery operation for 
added convenience.

Please read this manual carefully before using this product to ensure correct and safe usage: 

• 

The content of this manual is subject to change without notice 

• 

The appearance of the actual product can differ from the models described in this manual 

• 

Not all equipment mentioned in this manual has to be part of the set you have purchased 

• 

All optics are anti-fungus treated and anti-reflection coated for maximum light throughput 

2

Summary of Contents for 78904-00

Page 1: ...Installation and Operating Instructions MSU 200 Series Compound Microscope Model 78904 00...

Page 2: ...___________________________________________11 The field K hler diaphragm_____________________________________________________________________11 Adjusting the aperture diaphragm _______________________...

Page 3: ...cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before use and...

Page 4: ...sed disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination stateme...

Page 5: ...between 5 to 40 C and humidity should be within 20 to 80 RH Although the system is anti mold treated installing this product in a hot humid location may still result in the formation of mold or conde...

Page 6: ...and the slide The Numeric Aperture N A of the objective is an indication for the resolving power of the objective The total magnification can be calculated by multiplying the magnification of the eye...

Page 7: ...Safety device J Revolving nosepiece for 4 objectives C Tension adjustment K Objectives D Coaxial coarse and fine adjustment L Object stages mechanical stage or clips E On Off switch not visible M Cond...

Page 8: ...cord immediately and dry the microscope Assembling steps We always try to keep the number of assembly steps as low as possible but in some cases there are some steps to be taken The steps mentioned b...

Page 9: ...ied by Cole Parmer If it is lost or damaged choose one with the same specifications Replacement of the Power Cord If the power cord requires replacement ensure that it is suitably rated If in doubt co...

Page 10: ...to heavy according to your own preference Please note that when the specimen leaves the focus plane after focusing or the stage declines out of its own then you need to adjust the tension To tighten...

Page 11: ...ecimen 3 Rotate the field diaphragm adjustment ring to put the field diaphragm to the smallest position 4 Adjust the condenser to get the sharpest image 5 Adjust the center adjustment screw and center...

Page 12: ...per that is moistened with a drop of isopropanol 13 Clean the slide after use as well Caution Never put a drop of xylol or alcohol directly on the lens of the objective It may enter the objective and...

Page 13: ...ve the fuse compartment To do so insert a flat head screwdriver between the metal power tines and gently pry the fuse compartment loose with a slight downward and outward motion Insert the new fuse in...