Locking the eyepieces
To lock the eyepieces in binocular models please find the screw as indicated in picture (A). Please note that location
can be slightly rotated from model to model.
Connecting the power adapter
The MSU-100 Series supports a wide range of operating voltages: from 100 to 240 VAC Please use a grounded power
connection
1.
Make sure the power switch is off before connecting
2.
Insert the connector of power cord into the MSU-100 Series power socket, and make sure it connects well
3.
Insert the other connector of the adapter into the mains socket, and make sure it connects well
Do not bend or twist the power cord, it will be damaged
. Use the power cord that is supplied by Cole-Parmer. If it is
lost or damaged, choose one with the same specifications
Replacement of the Power Cord
If the power cord requires replacement, ensure that it is suitably rated. If in doubt consult your local dealer or qualified
electrician
Operation
Setting up the illumination:
For optimal contrast and resolution one should follow the procedure below:
A.
Place a specimen on the object stage and focus, using the 4x objective, with a fully opened iris diaphragm
B.
Turn light intensity to lowest position, then look through the eyepiece(s) and turn up to the comfortable light
intensity level
C.
Turn the condenser in the highest position
D.
Close the iris diaphragm, until it is just visible on the edge of the field of view
The microscope is properly set for use with the 4x objective. For each other magnification in bright field use, this
procedure should be repeated to ensure the best balance between contrast and resolution