Locking the eyepieces 

To lock the eyepieces in binocular models please find the screw as indicated in picture (A). Please note that location 
can be slightly rotated from model to model. 

Connecting the power adapter  

The MSU-100 Series supports a wide range of operating voltages: from 100 to 240 VAC Please use a grounded power 
connection 

1. 

Make sure the power switch is off before connecting 

2. 

Insert the connector of power cord into the MSU-100 Series power socket, and make sure it connects well 

3. 

Insert the other connector of the adapter into the mains socket, and make sure it connects well

Do not bend or twist the power cord, it will be damaged

. Use the power cord that is supplied by Cole-Parmer. If it is 

lost or damaged, choose one with the same specifications

Replacement of the Power Cord  

If the power cord requires replacement, ensure that it is suitably rated. If in doubt consult your local dealer or qualified 
electrician

Operation 

Setting up the illumination: 

For optimal contrast and resolution one should follow the procedure below: 

A. 

Place a specimen on the object stage and focus, using the 4x objective, with a fully opened iris diaphragm 

B. 

Turn light intensity to lowest position, then look through the eyepiece(s) and turn up to the comfortable light 
intensity level 

C. 

Turn the condenser in the highest position 

D. 

Close the iris diaphragm, until it is just visible on the edge of the field of view

The microscope is properly set for use with the 4x objective. For each other magnification in bright field use, this 
procedure should be repeated to ensure the best balance between contrast and resolution

Summary of Contents for 78903-90

Page 1: ...Installation and Operating Instructions MSU 100 Series Compound Microscope Model 78903 90...

Page 2: ...omponents of the microscope____________________________________________________ 7 Preparing the microscope for use __________________________________________________ 8 Assembling Steps _______________...

Page 3: ...a cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before use a...

Page 4: ...based disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination state...

Page 5: ...be between 5 to 40 C and humidity should be within 20 to 80 RH Although the system is anti mold treated installing this product in a hot humid location may still result in the formation of mold or con...

Page 6: ...and the slide The Numeric Aperture N A of the objective is an indication for the resolving power of the objective The total magnification can be calculated by multiplying the magnification of the eye...

Page 7: ...for 3 or 4 objectives B Safety device J Objectives C Tension adjustment K Object stages mechanical stage or clips D On Off switch not visible L Height adjustment lever condenser E Light intensity con...

Page 8: ...r head please pull out the power cord immediately and dry the microscope Assembling Steps Mounting the objectives 1 Rotate the coarse focusing knob to lower the stage to its lowest position 2 Install...

Page 9: ...is lost or damaged choose one with the same specifications Replacement of the Power Cord If the power cord requires replacement ensure that it is suitably rated If in doubt consult your local dealer o...

Page 10: ...ng with a S100x objective you need to lock the slide protection screw D The slide protection screw protects the slide by limiting the travel range of the mechanical stage This way the objectives will...

Page 11: ...h the object stage an Abbe condenser N A 1 25 is mounted The condenser can be adjusted in height by moving the lever beneath the mechanical stage N By adjusting the condenser you can focus the light o...

Page 12: ...er Clean the front lens of the S100x objective 11 Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a drop of isopropanol 12 Clean the slide after us...

Page 13: ...ust can be removed with high pressured air There will never be dust in the objectives if the objectives are not removed from the revolving nosepiece Caution Cleaning cloths containing plastic fibers c...

Page 14: ...as bulbs batteries fuses cords optical components or any add on accessories such as mechanical specimen holders that are not built into the microscope stage as an integral part of the original manufa...

Page 15: ...______________________________________________________ ___________________________________________________________________________________________________ _____________________________________________...

Page 16: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 100_Series_Compound Microscope_Manual...

Reviews: